The Recall Desk

FDA (Devices) recall action 92720

Philips North America recalled 3 products on 2023-08-16

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2023-08-16
Critical
0
Units
1,920

Why these products were recalled

The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2023-08-16

    Philips MR 7700 Magnetic Resonance System Smoke Detector Interlock Bypass

    Philips MR 7700 MRI systems can have their smoke detector interlock safety mechanism bypassed when power cycled after smoke detection, allowing continued operation despite potential fire hazards.

    Product
    MR 7700 Magnetic Resonance (MR) system, Product Numbers 782120 and 782130 (Upgrade to MR 7700)
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-08-16

    MR Imaging System Smoke Detector Interlock Vulnerable to Bypass via Power Cycling

    Philips Ingenia Elition X MR systems can have their smoke detection safety interlock bypassed by power cycling the equipment. This vulnerability could allow unsafe operation during hazardous conditions.

    Product
    Ingenia Elition X Magnetic Resonance (MR) system, Product Numbers 781358, 782107, and 782118 (SmartPath to Ingenia Elition X)
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-08-16

    Ingenia Elition S MR System smoke detection interlock vulnerable to operator bypass

    Philips Ingenia Elition S MR systems can be circumvented by power cycling to bypass the smoke detection safety interlock that locks after smoke is detected. This could allow continued operation despite fire hazards.

    Product
    Ingenia Elition S Magnetic Resonance (MR) system, Product Numbers 781357 and 782106
    Category
    Distribution
    0 states