The Recall Desk

FDA (Devices) recall action 92527

Datascope Corp. recalled 9 products on 2023-08-02

The FDA (Devices) publishes a recall like this one as 9 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 9 pages.

Products
9
Announced
2023-08-02
Critical
9
Units
2,894,481

Why these products were recalled

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–9 of 9

  • CriticalFDA (Devices)··2023-08-02

    MEGA Intra-Aortic Balloon Catheter Introducer Fracture Poses Arterial Injury Risk

    Datascope's MEGA 8Fr. intra-aortic balloon catheter introducer may fracture during insertion, leaving fragments that can damage blood vessels. The firm has reported 10 complaints, including 1 death.

    Product
    MEGA 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number MEGA 8Fr. 50cc IAB WITH STATLOCK…
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2023-08-02

    LINEAR intra-aortic balloon catheter introducer dilator fracture risk

    Datascope's LINEAR intra-aortic balloon catheters may develop a fractured introducer dilator that can injure blood vessels or cause dangerous blockages. The company has received 10 complaints including 1 death and 3 serious adverse events.

    Product
    LINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number LINEAR 7.5Fr. 34cc IAB WITH…
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2023-08-02

    Intra-Aortic Balloon Catheter Introducer Dilator Fracture During Insertion

    The YAMATO PLUS intra-aortic balloon catheter's introducer dilator may fracture during insertion, risking arterial damage, aortic damage, or blood clots. Datascope has received 10 complaints, including 1 death and 3 serious adverse events.

    Product
    YAMATO PLUS IAB; intra-aortic balloon catheter and accessories used to provide counter pulsation therapy in the aorta. Product Description Part Number YAMATO PLUS-R 30cc IAB WITH…
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2023-08-02

    Datascope SENSATION 7 Intra-Aortic Balloon Catheters Recalled for Introducer Fracture Risk

    The SENSATION 7 Fr. intra-aortic balloon catheter introducer dilator may fracture during insertion, potentially damaging blood vessels and requiring surgery. One death has been reported among 10 total complaints.

    Product
    SENSATION 7 Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number SENSATION 7 Fr. 34cc IAB WITH ACCESSORIES…
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2023-08-02

    Intra-Aortic Balloon Catheter Introducer Set fracture causes vascular injury

    Datascope Corp. recalls Reinforced Introducer Sets because the dilator may fracture during catheter insertion, risking arterial damage and embolization. One patient death has been reported.

    Product
    REINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number REINFORCED INTRODUCER…
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2023-08-02

    Datascope SENSATION PLUS Intra-Aortic Balloon Catheter Dilator Fracture Risk

    Datascope's SENSATION PLUS intra-aortic balloon catheter dilator may fracture during insertion, potentially damaging the femoral artery or aorta. The firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

    Product
    SENSATION PLUS 7.5Fr IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number SENSATION PLUS 7.5Fr.…
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2023-08-02

    Intra-Aortic Balloon Catheter kits recalled due to dilator fracture risk

    Datascope's MEGA 7.5Fr. intra-aortic balloon catheter kits may have introducers that fracture during insertion, risking arterial damage and embolization. The firm has received 10 complaints including 3 serious events and 1 death.

    Product
    MEGA 7.5Fr.IAB , Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number MEGA 7.5Fr. 40cc IAB WITH…
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2023-08-02

    Intra-Aortic Balloon Catheter Introducer Dilator May Fracture During Insertion

    Datascope TRANS-RAY IAB catheters may have introducer dilators that fracture during insertion, risking arterial damage or blood clots. The firm reports 1 death and 3 serious injuries.

    Product
    TRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number 7FR. 34cc TRANS-RAY IAB WITH ACCESSORIES D684-00-0545-01 TRANS-RAY 7FR.…
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2023-08-02

    Intra-aortic balloon catheter recall due to introducer dilator fracture risk

    Datascope Corp. is recalling SENSATION PLUS 8Fr. intra-aortic balloon catheters because the introducer dilator may fracture during insertion, risking blood vessel damage and surgical intervention. The company received 10 complaints, including 1 death.

    Product
    SENSATION PLUS 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number SENSATION PLUS 8Fr. 50cc…
    Category
    Distribution
    0 states