FDA (Devices) recall action 92501

Datascope Corp. recalled 2 products on 2023-09-06

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2023-09-06
Critical
0
Units
18,350

Why these products were recalled

Users were reporting that the device was not charging as expected. It was discovered that users were unaware that the Cardiosave console was not completely inserted into the hospital cart. If the console is not fully inserted back into the cart the battery(ies) will not charge.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • SevereFDA (Devices)··2023-09-06

    Cardiosave Rescue Intra-Aortic Balloon Pumps Fail to Charge When Improperly Inserted

    Datascope Corp. is recalling 9,175 units of Cardiosave Rescue IABP devices because batteries will not charge if the console is not fully inserted into the hospital cart. Users were often unaware of this insertion requirement.

    Product
    Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-09-06

    Cardiosave Hybrid Intra-Aortic Balloon Pump charging issue recall

    The Cardiosave Hybrid Intra-Aortic Balloon Pump console fails to charge if not fully inserted into the hospital cart. Users unaware of this requirement may attempt to use a non-charged device during critical cardiac care.

    Product
    Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53…
    Category
    Distribution
    0 states