FDA (Devices) recall action 92491

Datascope Corp. recalled 2 products on 2023-09-06

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2023-09-06
Critical
0
Units
18,350

Why these products were recalled

Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for these alarms that was not listed in the IFU: Patient Movement (coughing, general movement, and swallowing). Blood in the circuit, kins and abrasions of the tubing/catheter and certain patient conditions could also trigger these alarms. The HHE also identified hardware issues related to these alarms.

Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for these alarms that was not listed in the IFU: Patient Movement (coughing, general movement, and swallowing). Blood in the circuit, kinks/abrasions of the tubing/catheter and certain patient conditions could also trigger these alarms. The firm's HHE also identified hardware issues related to these alarms.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • SevereFDA (Devices)··2023-09-06

    Intra-Aortic Balloon Pumps recalled for unexpected gas loss and gain alarms

    Datascope's Cardiosave Hybrid Intra-Aortic Balloon Pumps may trigger unexpected alarms indicating gas loss or gain due to undocumented causes including patient movement and blood in the circuit. The FDA Class I recall affects 9,175 units.

    Product
    Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53…
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-09-06

    Cardiosave Rescue IABP Alarms May Trigger Unexpectedly During Therapy

    Datascope Corp. is recalling approximately 9,175 Cardiosave Rescue IABPs because unexpected gas-loss and gas-gain alarms may occur during therapy due to patient movement, blood in the circuit, and hardware issues.

    Product
    Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
    Category
    Distribution
    0 states
Datascope Corp. recalled 2 products on 2023-09-06 · The Recall Desk