The Recall Desk

FDA (Devices) recall action 92451

Luminex Corporation recalled 2 products on 2023-07-05

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2023-07-05
Critical
0
Units
4,101

Why these products were recalled

It is possible that in rare instances the VERIGENE CDF Nucleic Acid Test Kit (Part No. 20-005-022) and the VERIGENE Enteric Pathogens Nucleic Acid Test Kit (Part No. 20-005-023) could provide a false-negative result due to hydrophobic characteristics in specific FLOQSwab lots.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2023-07-05

    VERIGENE Clostridium difficile Nucleic Acid Test May Give False-Negative Results

    Luminex is recalling the VERIGENE C. difficile diagnostic test due to the possibility of false-negative results in rare instances caused by hydrophobic characteristics in specific FLOQSwab lots.

    Product
    VERIGENE Clostridium difficile Nucleic Acid Test, REF 30-002-022
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-07-05

    Verigene nucleic acid test kits recalled for rare false-negative results

    Luminex Corporation recalls Verigene nucleic acid test kits that may rarely produce false-negative results due to hydrophobic properties in certain FLOQSwab lots. False-negative results could fail to detect enteric pathogens in patient samples.

    Product
    Verigene Enteric Pathogens Nucleic Acid Test (EP), REF 30-002-023
    Category
    Distribution
    0 states