TELEFLEX LLC recalled 123 products on 2023-07-05

The FDA (Devices) publishes a recall like this one as 123 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 123 pages.

Products
123
Announced
2023-07-05
Critical
10
Units
6,948,980

Why these products were recalled

Reports of disconnection of the 15mm connector from the endotracheal tube (ET tube) for the affected products.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

26–50 of 123

  • SevereFDA (Devices)··2023-07-05

    Preformed AGT Endotracheal Tubes Recalled for Connector Disconnection Risk

    Teleflex is recalling 4,769 endotracheal tubes due to reports of 15mm connector disconnection. This defect could interrupt breathing support during airway management procedures.

    Product
    Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100180035
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Flexi-Set Uncuffed Endotracheal Tubes recalled for connector disconnection risk

    TELEFLEX LLC recalls Flexi-Set Uncuffed Endotracheal Tube and Stylet Sets due to reports of the 15mm connector disconnecting from the tube. Approximately 16,400 units distributed nationwide are affected.

    Product
    Flexi-Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 506540
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Preformed AGT Oral Endotracheal Tubes recalled for connector disconnection risk

    TELEFLEX is recalling 2,874 Preformed AGT Oral Endotracheal Tubes nationwide due to reports of the 15mm connector disconnecting from the tube. Affected users should immediately discontinue use.

    Product
    Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100181030
    Category
    Distribution
    1 state
  • CriticalFDA (Devices)··2023-07-05

    Slick Set Uncuffed Endotracheal Tube Connector Disconnection Recall

    Teleflex LLC is recalling approximately 10,650 units of Slick Set Uncuffed Endotracheal Tubes due to reports of 15mm connector disconnection from the tube.

    Product
    Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170140
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    FDA Recalls Teleflex Endotracheal Tubes for Connector Disconnection

    The FDA is recalling approximately 302,652 Teleflex preformed oral endotracheal tubes nationwide due to reports of disconnection between the 15mm connector and the tube. Healthcare providers should immediately discontinue use of affected units.

    Product
    Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111781050
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Endotracheal Tube Recalled for 15mm Connector Disconnection Risk

    TELEFLEX LLC is recalling 9,070 Preformed AGT Oral Endotracheal Tubes nationwide due to reports of 15mm connector disconnection. This medical device is used for airway management in hospital settings.

    Product
    Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111780080
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Endotracheal Tube Connector Failure Recall: Teleflex Slick Set Nationwide

    Teleflex is recalling 510 units of Slick Set Endotracheal Tubes with reported connector disconnection issues. The defect may prevent proper airway management during critical medical use.

    Product
    Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 150035
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    FDA recalls Teleflex endotracheal tubes for connector disconnection

    Teleflex endotracheal tubes with high-volume, low-pressure cuffs are recalled due to reports of 15mm connector disconnection. The recall affects 13,810 units distributed nationwide.

    Product
    Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080075
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Slick Set Cuffed Endotracheal Tube Connector Disconnection Recall

    Teleflex is recalling Slick Set Cuffed Endotracheal Tubes because the 15mm connector may disconnect from the tube. The recall affects 1,460 units distributed nationwide, including Puerto Rico.

    Product
    Slick Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170090
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Preformed AGT Oral Endotracheal Tube Connector May Disconnect

    Teleflex LLC is recalling Preformed AGT Oral Endotracheal Tubes due to reported connector disconnections. The recall affects 24,752 units distributed nationwide, including Puerto Rico.

    Product
    Preformed AGT Oral Endotracheal Tube uncuffed/plain - Murphy, REF 100181055
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Endotracheal Tubes Recalled Due to Connector Disconnection Issues

    Teleflex LLC is recalling 17,102 Endotracheal Tubes (REF 100382040) following reports of disconnection of the 15mm connector from the tube. The affected units were distributed nationwide, including Puerto Rico.

    Product
    Endotracheal Tube oral/nasal uncuffed/plain - Murphy, REF 100382040
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Endotracheal Tube Connector Disconnection Recall Due to Device Malfunction

    Teleflex is recalling 20,042 endotracheal tubes due to reports of 15mm connector disconnection that could compromise airway management. The FDA classified this as a Class I recall.

    Product
    Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111780040
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Endotracheal Tube Connector Disconnection Risk in Flexi-Set Kit

    Flexi-Set Cuffed Endotracheal Tube kits are being recalled due to reports of the 15mm connector disconnecting from the tube. Affected units were distributed nationwide.

    Product
    Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504555
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Preformed AGT Oral Endotracheal Tubes Recalled for Connector Disconnection

    Teleflex is recalling 35,513 preformed AGT oral endotracheal tubes nationwide due to reported connector disconnection from the 15mm tube connector.

    Product
    Preformed AGT Oral Endotracheal Tube Murphy Eye, High Volume, Low Pressure Cuff, REF 111781070
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Endotracheal Tube Sets Recalled for Connector Disconnection Risk

    Teleflex is recalling Flexi-Set Cuffed Endotracheal Tube Sets due to reported 15mm connector disconnection. The recall affects 92,930 units distributed throughout the United States, including Puerto Rico.

    Product
    Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504565
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Endotracheal Tube Connector Disconnection Risk Prompts Medical Device Recall

    Teleflex is recalling Slick Set Uncuffed Endotracheal Tube and Stylet Sets due to reports that the 15mm connector may disconnect from the tube. The recall affects 447 units distributed nationwide.

    Product
    Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 150040
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Slick Set Endotracheal Tubes recalled for connector disconnection

    TELEFLEX is recalling 2,382 Slick Set Uncuffed Endotracheal Tubes due to reports of 15mm connector disconnection. Units were distributed nationwide including Puerto Rico.

    Product
    Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 150025
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Teleflex Flexi-Set Endotracheal Tubes Recalled for Connector Disconnection

    Teleflex is recalling 32,553 units of Flexi-Set Uncuffed Endotracheal Tubes nationwide due to reports of 15mm connector disconnections that could compromise airway management.

    Product
    Flexi-Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 506530
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    TELEFLEX Endotracheal Tubes Recalled for 15mm Connector Disconnection

    TELEFLEX is recalling 5,990 endotracheal tubes due to reports of 15mm connector disconnection. Affected units were distributed nationwide, including Puerto Rico.

    Product
    Endotracheal Tube oral/nasal uncuffed/plain - Magill, REF 100380030
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Teleflex endotracheal tubes recalled due to connector disconnection

    Teleflex has recalled 49,050 endotracheal tubes due to reported disconnection of the 15mm connector. Affected units were distributed nationwide, including Puerto Rico.

    Product
    Endotracheal Tube oral/nasal Murphy Eye, High Volume, Low Pressure Cuff, REF 112082055
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Teleflex Slick Set Endotracheal Tube Connector Disconnection Recall

    Teleflex is recalling 18,900 Slick Set Uncuffed Endotracheal Tubes due to reports of 15mm connector disconnection. The affected units were distributed nationwide in the United States and Puerto Rico.

    Product
    Slick Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 170125
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Flexi-Set Endotracheal Tube Recalled for Connector Disconnection

    Teleflex is recalling 18,353 Flexi-Set Uncuffed Endotracheal Tube sets nationwide due to reports of 15mm connector disconnection. The defect could affect ventilation during use.

    Product
    Flexi-Set Uncuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 506535
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Flexi-Set Endotracheal Tubes recalled for connector disconnection

    Teleflex is recalling over 1 million Flexi-Set Cuffed Endotracheal Tube kits due to reports of 15mm connector disconnection from the tube. The recall affects products distributed nationwide.

    Product
    Flexi-Set Cuffed Endotracheal Tube and Stylet Set, oral/nasal, REF 504570
    Category
    Distribution
    1 state
  • SevereFDA (Devices)··2023-07-05

    Endotracheal Tubes Recalled for Potential 15mm Connector Disconnection

    TELEFLEX is recalling 4,160 endotracheal tubes due to reports of the 15mm connector disconnecting from the tube. Affected units were distributed nationwide.

    Product
    Endotracheal Tube oral/nasal Magill, High Volume, Low Pressure Cuff, REF 112080060
    Category
    Distribution
    1 state