The Recall Desk

FDA (Devices) recall action 92321

Abiomed, Inc. recalled 6 products on 2023-07-26

The FDA (Devices) publishes a recall like this one as 6 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 6 pages.

Products
6
Announced
2023-07-26
Critical
0
Units
55,512

Why these products were recalled

There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–6 of 6

  • SevereFDA (Devices)··2023-07-26

    Impella CP cardiac pump recalls due to motor housing interaction with stents

    Abiomed is recalling Impella CP with SmartAssist blood pumps due to risk of motor housing interaction with transcatheter aortic valve replacement devices, which can damage the pump's impeller blades and reduce blood flow.

    Product
    Impella CP with SmartAssist intravascular micro axial blood pump, Product Numbers 0048-0024, 0048-0045, 1000080
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-07-26

    Impella LD cardiac pump recalled for impeller damage and embolization risk

    Abiomed recalls Impella LD blood pumps due to potential interaction with aortic valve replacement devices that can damage the pump impeller and cause blood clots. Patients with these devices should contact their healthcare provider.

    Product
    Impella LD intravascular micro axial blood pump, Product Number 005082
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-07-26

    Impella 5.0 Blood Pump Recalled for Potential Impeller Blade Destruction Risk

    Abiomed's Impella 5.0 intravascular blood pump is recalled due to risk of impeller blade destruction when interacting with heart valve replacements. The damaged pump can cause reduced blood flow and potentially release fractured material into the bloodstream.

    Product
    Impella 5.0 intravascular micro axial blood pump, Product Number 005062
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-07-26

    Abiomed Impella 5.5 pump recalled due to impeller blade damage risk

    Abiomed is recalling the Impella 5.5 with SmartAssist pump due to a risk of impeller blade destruction when used with aortic valve replacement devices. Damage can result in low blood flow and systemic embolization, where broken pump pieces may enter the bloodstream.

    Product
    Impella 5.5 with SmartAssist intravascular micro axial blood pump, Product Numbers 0550-0008 and 1000100
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-07-26

    Impella 2.5 blood pump recalled due to impeller blade damage risk

    Impella 2.5 intravascular blood pumps may experience unintended interaction with aortic valve replacement stents, damaging impeller blades and reducing blood flow. Fragments may circulate through the bloodstream.

    Product
    Impella 2.5 intravascular micro axial blood pump, Product Number 005042
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-07-26

    Impella CP Blood Pump Recalled for Blade Destruction From Stent Interaction

    Abiomed's Impella CP cardiac pump can be damaged by interaction with replacement heart valves, potentially causing blood clots. The FDA classified this as a Class I recall of 9,252 units worldwide.

    Product
    Impella CP intravascular micro axial blood pump, Product Number 0048-0032
    Category
    Distribution
    0 states