Vyaire Medical recalled 2 products on 2023-06-21

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2023-06-21
Critical
0
Units
1,302

Why these products were recalled

Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops responding (APP Hang). When this occurs, if the user continues to interact with the touchscreen, a message will appear that states DeviceSoftware.Application is not responding or a decommission screen will appear on the user interface. The ventilator will issue both an audible and visual alarm. Ventilation continues without interruption with the settings applied prior to the…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2023-06-21

    bellavista 1000e Ventilator Software May Hang, Affecting Device Interface

    Vyaire Medical is recalling certain bellavista 1000e ventilators due to software issues that may cause the touchscreen to become unresponsive. The recall also addresses potential operator error risks if ventilator settings are not confirmed before application.

    Product
    bellavista 1000e Ventilator, Catalogue Number 301.100.130; Continuous Use Ventilator
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-06-21

    bellavista 1000 Ventilator Recalled for Software Hang and Use Error Risk

    Vyaire Medical is recalling approximately 910 units of the bellavista 1000 Ventilator due to a software issue that can cause the touchscreen to become unresponsive and a potential use error risk if operators do not confirm settings are appropriate for the patient.

    Product
    bellavista 1000 Ventilator, Catalogue Number 301.100.030; Continuous Use Ventilator
    Category
    Distribution
    0 states