The Recall Desk

FDA (Devices) recall action 92230

NRT X-RAY A/S recalled 2 products on 2023-06-07

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2023-06-07
Critical
0
Units
45

Why these products were recalled

There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2023-06-07

    Medical Device Recall: X-Ray Detector Drive Shaft May Break

    The FDA is recalling 32 X-ray detector units due to a potential drive shaft failure that could cause uncontrolled detector movement. Affected units are distributed in 10 US states.

    Product
    Intelli-C, Right side suspended Tabletop, REF: 03400000
    Category
    Distribution
    10 states
  • HighFDA (Devices)··2023-06-07

    X-Ray Machine Detector May Move Unexpectedly Due to Drive Shaft Failure

    FDA recalls the Intelli-C X-Ray machine's left side suspended tabletop model due to potential drive shaft failure in the detector lift mechanism, which could cause uncontrolled detector movement.

    Product
    Intelli-C, Left side suspended Tabletop, REF: 03400010
    Category
    Distribution
    10 states