The Recall Desk

FDA (Devices) recall action 92202

Philips North America recalled 4 products on 2023-08-09

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2023-08-09
Critical
0
Units
79

Why these products were recalled

Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2023-08-09

    CT Scanner Metal Mounting Box May Become Expelled, Posing Injury Risk

    Philips is recalling Incisive CT Plus imaging systems due to a metal mounting box that may separate from the rotating scanner and potentially injure users.

    Product
    Incisive CT Plus-Computed Tomography X-Ray System Model: 728149
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-08-09

    Philips CT scanner heat exchanger box may detach and cause injury

    The metal mounting box on the rotating scanner of the Incisive CT system may become detached and contact other components, potentially injuring users or causing system shutdown.

    Product
    Incisive CT-Computed Tomography X-Ray System Model: 728143
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-08-09

    Incisive CT Computed Tomography System: Metal Mounting Box May Detach

    A metal mounting box in the Incisive CT scanner may become expelled, potentially injuring users or causing the system to shut down. Philips is recalling 20 affected units distributed internationally.

    Product
    Incisive CT for Brazil SKD-Computed Tomography X-Ray System Model: 728146
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-08-09

    Philips Incisive CT Scanner Mounting Box Poses Injury Risk

    A metal mounting box in the Incisive CT Power scanner may detach and potentially injure operators or cause system failure. Philips is recalling 53 affected units distributed worldwide.

    Product
    Incisive CT Power (China)-Computed Tomography X-Ray System Model: 728148
    Category
    Distribution
    0 states