The Recall Desk

FDA (Devices) recall action 92056

Beckman Coulter, Inc. recalled 4 products on 2023-05-24

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2023-05-24
Critical
0
Units
1,836

Why these products were recalled

Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.

Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have been investigated.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2023-05-24

    Urine Microscopy Analyzer Software Defect May Prevent Carryover Detection

    Beckman Coulter is recalling the DxU 840m Iris Urine Microscopy Analyzer due to a software issue where a carryover detection flag may fail to display, potentially resulting in undetected sample contamination and incorrect test results.

    Product
    DxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic device, urine analysis
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-05-24

    Beckman Coulter Urine Analyzer Software Flag Display Issue Affects Carryover Detection

    Beckman Coulter is recalling 107 DxU 850m Iris Urine Microscopy Analyzers worldwide due to an intermittent software issue where a quality-control flag is enabled but not displayed, potentially preventing detection of sample carryover contamination.

    Product
    DxU 850m Iris Urine Microscopy Analyzer Catalog Number C49513, in-vitro diagnostic device, urine analysis
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-05-24

    Urine Microscopy Analyzer Software Display Defect May Prevent Contamination Detection

    Beckman Coulter's iQ200 Series urine analyzers may fail to display a flag indicating prior samples, potentially preventing detection of sample contamination from carryover events.

    Product
    iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Numbers C10683 and 700-3375; in-vitro diagnostic device, urine analysis, c) iQ200 Select 1500, Part Number 700-3347, d) iQ200 Sprint, Part Number 700-3325
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-05-24

    Urine Analysis Diagnostic Device Software Issue Prevents Carryover Detection

    Beckman Coulter is recalling UA WIN 10 diagnostic kits because an optional software flag for detecting sample carryover was enabled but not displayed, preventing proper investigation of potential test contamination.

    Product
    UA WIN 10 Kit, Catalog Number C52900, in-vitro diagnostic device, urine analysis
    Category
    Distribution
    0 states