Philips Respironics, Inc. recalled 4 products on 2023-06-14

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2023-06-14
Critical
1
Units
120,007

Why these products were recalled

Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider.

The Trilogy Evo Universal ventilator provides continuous or intermittent positive pressure ventilation for the care of individuals who require mechanical ventilation. Trilogy Evo Universal is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and pulse rate data when integrated with the appropriate accessories. The ventilator is suitable for use in institutional and hospital settings and non-emergency transport settings; for example, wheelchair. It may be used for both invasive and non-invasive ventila…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • SevereFDA (Devices)··2023-06-14

    Philips Respironics Ventilators Recalled for Flow Sensor Debris Accumulation

    Philips Respironics recalled approximately 57,395 ventilators due to environmental debris potentially accumulating on the flow sensor, which may prevent accurate delivery of pressure and volume to patients, risking hypoxemia.

    Product
    Trilogy Evo, Garbin Evo, and Aeris EVO, Model Numbers DS2110X11B, IN2110X15B, JP2110X16B, AU2110X15B, LA2110X15B, CA2110X12B, BR2110X18B, KR2110X15B, IA2110X15B, FP2110X10, FR2110X14B, ND2110X15B, IT2110X21B, ES2110X15B, DE2110X13B, BL2110X15B, GB2110X15B, EU2110X15B, EE2110X15B,
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-14

    Trilogy Evo O2 EV300 respiratory devices recalled for debris accumulation

    Philips Respironics recalled 18,855 Trilogy Evo O2 EV300 devices because environmental debris may accumulate on the flow sensor, potentially causing inaccurate delivery of pressure or flow. This could result in patient hypoxemia if not promptly addressed.

    Product
    Trilogy Evo O2 EV300, Model Numbers DS2200X11B, IN2200X15B, FX2200X15B, CA2200X12B, FR2200X14B, ND2200X15B, ES2200X15B, BL2200X15B, GB2200X15B, EU2200X15B, EE2200X15B, and TR2200X15B
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2023-06-14

    Trilogy Evo Universal Ventilator Class I Recall—22,237 Units Affected

    The FDA has issued a Class I recall for the Trilogy Evo Universal Ventilator Model DS2000X11B, affecting 22,237 units distributed nationwide and globally. The device is manufactured by Philips Respironics, Inc.

    Product
    Trilogy Evo Universal Ventilator, Model Number DS2000X11B
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2023-06-14

    Philips Trilogy and LifeVent ventilator models recalled for flow sensor debris

    Environmental debris accumulation may block flow sensors in Philips ventilators, potentially preventing accurate pressure and flow delivery to patients.

    Product
    Trilogy Evo O2 and LifeVent EVO2, Model Numbers DS2100X11B, IN2100X15B, IN2100X19, FX2100X15B, LA2100X15B, IA2100X15B, FP2100X10, FR2100X14B, ND2100X15B, IT2100X21B, ES2100X15B, DE2100X13B, EU2100X15B, EU2100X19, EE2100X15B, and SP2100X26B
    Category
    Distribution
    0 states