FDA (Devices) recall action 91938

Draeger Medical Systems, Inc. recalled 2 products on 2023-05-03

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2023-05-03
Critical
0
Units
26,232

Why these products were recalled

The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2023-05-03

    Draeger Infinity M300 Patient Monitor Software Displays Inaccurate ECG Readings

    Draeger Infinity M300/M300+ patient monitoring systems may display or print ECG waveforms with inaccurate QRS amplitudes due to software peaks being dropped on narrow waveforms.

    Product
    Draeger Infinity M300 and M300+, for use with the ICS to monitor ECG and pulse oximetry on ambulatory and non-ambulatory adult and pediatric patients using wireless communication over the Infinity patient monitoring network,
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-05-03

    Draeger Infinity CentralStation ECG waveform display inaccuracy due to software defect

    Draeger Infinity CentralStation patient monitors may display ECG waveform amplitudes lower than actual values due to a software bug affecting versions VG2.1.3 and lower.

    Product
    Draeger Infinity CentralStation (ICS), centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment, Software versions VG2.1.3 and lower
    Category
    Distribution
    0 states