The Recall Desk

FDA (Devices) recall action 91854

Bio-Rad Laboratories, Inc. recalled 2 products on 2023-04-19

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2023-04-19
Critical
0
Units
1,769

Why these products were recalled

Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8  96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1  98.8%). The reduction in specificity may lead to an increase in the number of false positives.

Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots. Internal testing at Bio-Rad has demonstrated that Lot No. 301481 exhibits a specificity of 94.5% (95% CI 91.8  96.4) whereas the IFU indicates a specificity of 97.8% (95% CI 96.1  98.8%). The reduction in specificity may lead to an increase in the number of false positivesBio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2023-04-19

    BioPlex 2200 Anti-CCP Reagent Pack Recalled for Reduced Diagnostic Specificity

    Bio-Rad is recalling BioPlex 2200 Anti-CCP Reagent Pack, Lot 301481, because the test's specificity is lower than claimed, which may increase false positive results. The diagnostic kit was distributed worldwide including the US, Canada, and several European countries.

    Product
    BioPlex 2200, Anti-CCP Reagent Pack, REF 665-3250, IVD
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-04-19

    BioPlex 2200 Anti-CCP Reagent Kit recalled due to reduced test specificity

    Bio-Rad is recalling BioPlex 2200 Anti-CCP reagent kits due to reduced specificity that may cause false positive results. Four customer complaints documented discrepant results during validation testing.

    Product
    BioPlex 2200, Anti-CCP Calibrator Set, REF 663-3200, IVD
    Category
    Distribution
    0 states