The Recall Desk

FDA (Devices) recall action 91652

Datascope Corp. recalled 4 products on 2023-03-29

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2023-03-29
Critical
0
Units
1,461

Why these products were recalled

There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2023-03-29

    Cardiosave IABP Pneumatic Interface Module Recalled for Performance Issue

    Datascope Corp. recalls the Cardiosave IABP Pneumatic Interface Module due to a Safety Disk performance issue that may reduce balloon inflation. Ten specific units with identified serial numbers are affected.

    Product
    Cardiosave Intra-Aortic Balloon Pump (IABP) Pneumatic Interface Module (PIM) , Model Number 0997-00-1178
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-03-29

    Cardiosave Hybrid Intra-Aortic Balloon Pump Performance Issue Recall

    Datascope Corp. is recalling specific Cardiosave Hybrid intra-aortic balloon pumps due to a Safety Disk performance issue that may reduce balloon inflation and therapeutic benefit.

    Product
    Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-03-29

    Cardiosave IABP Safety Disk Model 0202-00-0140 Recalled for Performance Defect

    Datascope Corp. is recalling 669 Cardiosave IABP Safety Disks due to a performance defect that may reduce balloon inflation during cardiac support therapy. The affected devices may not provide adequate augmentation to patients requiring mechanical circulatory support.

    Product
    Cardiosave Intra-Aortic Balloon Pump (IABP) Safety Disk, Model Number 0202-00-0140
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-03-29

    Cardiosave Hybrid intra-aortic balloon pump safety disk defect may reduce therapy efficacy

    Datascope's Cardiosave Hybrid intra-aortic balloon pump may have a faulty Safety Disk that reduces balloon inflation and cardiac augmentation. No injuries reported.

    Product
    Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, and 0998-00-0800-65
    Category
    Distribution
    0 states