St. Jude Medical recalled 4 products on 2023-03-15

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2023-03-15
Critical
0
Units
40,187

Why these products were recalled

Radiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the potential to cause interference with other medical devices, such as implantable neurostimulators, pacemakers, cardiac defibrillators, continuous glucose monitors, or other bio-wearable sensors when in close proximity to an active Hospital or Patient Electronics System (i.e., during readings).

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2023-03-15

    CardioMEMS Patient Electronics System Recalled for Excessive Radiofrequency Emissions

    CardioMEMS heart-failure monitoring devices are recalled due to radiofrequency emissions that can interfere with other medical devices like pacemakers when operating nearby.

    Product
    CardioMEMS HF System Patient Electronics System (PES), Model number CM1100, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-03-15

    CardioMEMS Heart Monitoring System Emissions May Interfere With Pacemakers

    FDA recalls CardioMEMS HF System patient monitor for radiofrequency signal emissions that can interfere with pacemakers, defibrillators, and other implanted medical devices. The emissions exceed radiofrequency limits.

    Product
    CardioMEMS HF System Patient Electronics System (PES), Model number CM1010, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-03-15

    CardioMEMS Hospital Electronics System RF Emissions May Interfere With Implanted Devices

    St. Jude Medical is recalling the CardioMEMS HF System (Model CM3000) due to excessive radiofrequency emissions that may interfere with pacemakers, defibrillators, and other implanted devices during use. The recall affects 1,905 units distributed worldwide.

    Product
    CardioMEMS HF System Hospital Electronics System (HES), Model number CM3000, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-03-15

    CardioMEMS Heart Failure Monitor Recalled for Radiofrequency Emissions

    St. Jude Medical is recalling certain CardioMEMS heart failure monitoring devices due to excessive radiofrequency emissions that could interfere with pacemakers, defibrillators, and other implanted medical devices when in close proximity.

    Product
    CardioMEMS HF System Patient Electronics System (PES), Model number CM1000, The CardioMEMS HF System provides pulmonary artery (PA) hemodynamic data used for the monitoring and management of heart failure (HF) patients. The system measures changes in PA pressure which…
    Category
    Distribution
    0 states