Datascope Corp. recalled 2 products on 2023-03-15

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2023-03-15
Critical
1
Units
—

Why these products were recalled

An unexpected shutdown of the IABP may occur due to loss of communication between the Executive Processor PCBA and the Video Generator PCBA. This may result in an interruption to therapy which may threaten the hemodynamic stability of the supported patient as the user is left unaware to the status of the Cardiosave IABP.

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–2 of 2

  • SevereFDA (Devices)··2023-03-15

    Cardiosave Hybrid IABP Recalled for Risk of Unexpected Shutdown

    The Cardiosave Hybrid intra-aortic balloon pump (IABP) is being recalled because it may unexpectedly shut down due to communication loss between internal components. The shutdown could interrupt therapy in cardiac patients and compromise their hemodynamic stability.

    Product
    Cardiosave Hybrid Model Nos. 0998-00-0800-31 0998-00-0800-32 0998-00-0800-33 0998-00-0800-34 0998-00-0800-35 0998-00-0800-45 0998-00-0800-52 0998-00-0800-53 0998-00-0800-55 0998-00-0800-65 0998-UC-0800-31* 0998-UC-0800-33* 0998-UC-0800-52* 0998-UC-0800-53*…
    Category
    Distribution
    45 states
  • CriticalFDA (Devices)··2023-03-15

    Cardiosave Rescue intra-aortic balloon pump may shut down unexpectedly

    The Cardiosave Rescue intra-aortic balloon pump may suddenly stop working due to communication failure between internal circuit boards, interrupting patient therapy and threatening hemodynamic stability.

    Product
    Cardiosave Rescue Model Nos. 0998-00-0800-75 0998-00-0800-83 0998-00-0800-85