The Recall Desk

FDA (Devices) recall action 91564

Unetixs Vascular, Inc. recalled 4 products on 2023-03-08

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2023-03-08
Critical
0
Units
7,000

Why these products were recalled

The current luer fittings and adaptors allow for the possibility of the air line to be misconnected to an intravenous line, which may cause an air embolism.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2023-03-08

    Vascular diagnostic device recalled for air line misconnection risk

    Unetixs Vascular is recalling the MultiLab Series II REVO vascular diagnostic device because faulty connectors may allow the air line to be misconnected to an IV line, potentially causing a dangerous air embolism.

    Product
    MultiLab Series II REVO non-invasive vascular diagnostic device, Model Number 12950-0000-01
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-03-08

    Medical Device Recall: Vascular Diagnostic Device Air Embolism Risk

    Unetixs Vascular is recalling 1706 units of the MultiLab Series II 2CP & 2CP Express vascular diagnostic device due to luer fittings that could allow air lines to be misconnected to intravenous lines, potentially causing air embolism.

    Product
    MultiLab Series II 2CP & 2CP Express non-invasive vascular diagnostic device, Model Number 11996-0000-01
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-03-08

    Vascular diagnostic devices recalled due to air embolism misconnection risk

    Unetixs Vascular recalls MultiLab Series II vascular diagnostic devices worldwide due to a design flaw where air lines can be misconnected to IV lines, potentially causing air embolism.

    Product
    MultiLab Series II LHS TI & ERGO non-invasive vascular diagnostic device, Model Number 11949-0000-01
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2023-03-08

    Vascular Diagnostic Device Recalled for Risk of Air Line Misconnection

    Unetixs Vascular recalls MultiLab Series II ROODRA vascular diagnostic devices due to luer fittings that may allow air line misconnection to intravenous lines, potentially causing air embolism.

    Product
    MultiLab Series II ROODRA non-invasive vascular diagnostic device, Model Number 11986-0000-01
    Category
    Distribution
    0 states