The Recall Desk

FDA (Devices) recall action 91290

Johnson & Johnson Surgical Vision, Inc. recalled 3 products on 2023-02-01

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2023-02-01
Critical
0
Units
11,161

Why these products were recalled

Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2023-02-01

    Veritas Advanced Infusion Packs Recalled Due to Weld Gap Defect

    Johnson & Johnson Surgical Vision is recalling Veritas Advanced Infusion Packs for a weld defect that may impair vacuum delivery during ophthalmic surgery. This could result in surgical delays and transient corneal swelling.

    Product
    Veritas Advanced Infusion Packs, REF: VRT-AI
    Category
    Distribution
    8 states
  • HighFDA (Devices)··2023-02-01

    FUSION Dual Pump Packs recalled for weld defect affecting surgery

    FUSION Dual Pump Packs are being recalled for weld defects that may impair vacuum delivery during eye surgery, risking surgical delays and corneal swelling. No injuries have been reported.

    Product
    FUSION Dual Pump Packs, REF: OPO73
    Category
    Distribution
    8 states