The Recall Desk

FDA (Devices) recall action 91139

GE Medical Systems, LLC recalled 8 products on 2022-12-28

The FDA (Devices) publishes a recall like this one as 8 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 8 pages.

Products
8
Announced
2022-12-28
Critical
0
Units
23,805

Why these products were recalled

Potential for batteries that have not been replaced at 2 years, as recommended in the Service Manual, they can fail or possibly emit smoke, or catch fire.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–8 of 8

  • HighFDA (Devices)··2022-12-28

    Ultrasound System Batteries May Fail, Emit Smoke, or Catch Fire

    GE Medical Systems is recalling Vivid q N ultrasound systems due to potential battery failures that could emit smoke or catch fire if batteries are not replaced per the 2-year service manual recommendation.

    Product
    Vivid q N ultrasound. Not marketed in the US. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skelet
    Category
    Distribution
    52 states
  • HighFDA (Devices)··2022-12-28

    Vivid i N ultrasound device batteries may fail, smoke, or catch fire

    GE Medical Systems recalls Vivid i N ultrasound machines worldwide because unreplaced batteries may fail, emit smoke, or catch fire. Users should replace batteries every 2 years per the service manual.

    Product
    Vivid i N ultrasound. Not marketed in the US. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal
    Category
    Distribution
    52 states
  • HighFDA (Devices)··2022-12-28

    GE Vivid S6 N Ultrasound Battery Fire Risk Recall

    GE Vivid S6 N ultrasound systems may experience battery failure, smoke, or fire if batteries are not replaced every 2 years as recommended. Clinicians should replace batteries per manufacturer Service Manual instructions.

    Product
    Vivid S6 N ultrasound. Not marketed in the US. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal
    Category
    Distribution
    52 states
  • HighFDA (Devices)··2022-12-28

    GE Vivid S6 Ultrasound Systems Recalled for Battery Fire Risk

    GE Vivid S6 ultrasound systems are recalled for a potential battery hazard. If not replaced at the recommended 2-year interval, batteries may fail, emit smoke, or catch fire.

    Product
    Vivid S6 ultrasound. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superf
    Category
    Distribution
    52 states
  • HighFDA (Devices)··2022-12-28

    Vivid S5 N Ultrasound Systems Recalled for Potential Battery Fire Hazard

    GE Medical Systems recalled 1,400 Vivid S5 N ultrasound systems for batteries that may fail, emit smoke, or catch fire if not replaced every 2 years as recommended.

    Product
    Vivid S5 N ultrasound. Not marketed in the US. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal
    Category
    Distribution
    52 states
  • HighFDA (Devices)··2022-12-28

    GE Vivid S5 Ultrasound Systems Battery Failure Fire Risk

    GE Medical Systems is recalling Vivid S5 ultrasound systems because batteries may fail if not replaced at the recommended 2-year intervals, potentially causing smoke or fire.

    Product
    Vivid S5 Ultrasound. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superf
    Category
    Distribution
    52 states
  • HighFDA (Devices)··2022-12-28

    GE Vivid i Ultrasound Systems Battery Failure Recall

    GE Medical Systems is recalling Vivid i ultrasound systems because batteries not replaced every two years may fail, emit smoke, or catch fire. About 7,931 units were distributed worldwide.

    Product
    Vivid i ultrasound. Marketed under the following 510(k) numbers: K033139, K061525, K082374, K092140, K102388, K121062). Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cepha
    Category
    Distribution
    52 states
  • HighFDA (Devices)··2022-12-28

    GE Vivid q Ultrasound Systems: Battery Fire Risk Recall

    GE is recalling Vivid q ultrasound systems because batteries not replaced every 2 years as recommended may fail and emit smoke or catch fire.

    Product
    Vivid q ultrasound. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superfi
    Category
    Distribution
    52 states