The Recall Desk

FDA (Devices) recall action 91097

Randox Laboratories Ltd. recalled 2 products on 2022-12-14

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2022-12-14
Critical
0
Units
40

Why these products were recalled

Update to the carryover avoidance technical bulletin to detail that IgA and IgM assays should not be run immediately after Fructosamine Catalogue Number FR3133 or FR4030 on the RX instruments testing order. Can lead to an elevation in Quality Control and Patient results of up to +13% for lgA and +51% for lgM respectively, which may lead to a delay in running patient samples or erroneous elevated test results

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2022-12-14

    IgM and IgA Assay Kits Recalled for Test Carryover Risk

    Randox Laboratories recalls IgM and IgA immunoturbidimetric assay kits due to carryover contamination that can cause elevated test results when run immediately after Fructosamine tests on RX instruments.

    Product
    IgM Immunoturbidimetric Assay-In vitro determination of IgA and IgM in serum and plasma Catalog Number: IM3834
    Category
    Distribution
    4 states
  • ModerateFDA (Devices)··2022-12-14

    IgA and IgM Immunoturbidimetric Assay Testing Order Carryover Avoidance Update

    Randox Laboratories updated its carryover avoidance protocol for IgA and IgM immunoturbidimetric assays. Running these assays immediately after Fructosamine tests on RX instruments can cause falsely elevated results (up to +13% for IgA, +51% for IgM).

    Product
    IgA Immunoturbidimetric Assay- In vitro determination of IgA and IgM in serum and plasma Catalog Number: IA3832
    Category
    Distribution
    4 states