Greer Laboratories, Inc. recalled 2 products on 2022-12-07

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2022-12-07
Critical
0
Units
320

Why these products were recalled

The product was recalled because stability evaluation of the lots indicated a decrease in reactivity that could lead to a potential false-negative result interpretation. The issue was identified by the firm after failing to meet lot requirements for potency. The lot release error was discovered on October 08, 2021, after a failed potency test was determined during 6-month stability evaluation of retention material for the lots. The subsequent manufacturing investigation, which included a review of the lot release potency documentation, found that the initial lot release potency test for the l…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2022-12-07

    Medical laboratory reagent recalled for potential false-negative test results

    Greer Laboratories recalled a hypersensitivity pneumonitis test reagent (Lot 389591) due to decreased reactivity that could cause false-negative results. The issue affects 218 units distributed across 12 U.S. states and two countries.

    Product
    GREER M. FAENI, 0.25 mL 20 MG/ML, Item: K6, Analyte Specific Reagent, HYPERSENSITIVITY PNEUMONITIS test system, gel diffusion kit for demonstrating precipitating antibodies.
    Category
    Distribution
    11 states
  • HighFDA (Devices)··2022-12-07

    Medical device test system recalled for decreased reactivity and false-negative results

    Greer Laboratories recalled a hypersensitivity pneumonitis test kit (102 units, Lot 389592) after stability testing revealed decreased reactivity that could cause false-negative results. The issue was discovered in October 2021.

    Product
    GREER DIA - KIT ANTIGEN M. FAENI, 2.0mL 20 MG/ML, Analyte Specific Reagent, Item: K62, HYPERSENSITIVITY PNEUMONITIS test system, gel diffusion kit for demonstrating precipitating antibodies.
    Category
    Distribution
    11 states