The Recall Desk

FDA (Devices) recall action 91070

Integrity Implants Inc. recalled 3 products on 2022-12-14

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2022-12-14
Critical
0
Units
4,298

Why these products were recalled

Potential for weld separation of the Upper Tulip Head component from the Lower Tulip Head component, which would result in the separation of the Tulip Head from the Screw Shank.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–3 of 3

  • HighFDA (Devices)··2022-12-14

    Spinal implant screws recalled for potential weld separation

    Integrity Implants recalls 1,369 spinal implant screws due to potential weld separation between components. No injuries have been reported.

    Product
    LineSider Spinal System 5.5/6.0mm, Non-mod MIS 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped individually, as…
    Category
    Distribution
    16 states
  • HighFDA (Devices)··2022-12-14

    LineSider Spinal System Screws Recalled for Potential Weld Separation

    Integrity Implants is recalling 1,702 LineSider Spinal System pedicle screws due to potential weld separation between screw head components that could result in device failure.

    Product
    LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped…
    Category
    Distribution
    16 states
  • HighFDA (Devices)··2022-12-14

    Spinal Implant Screws Recalled Due to Potential Weld Failure

    LineSider Spinal System screws are recalled due to potential weld separation between components that could cause the Tulip Head to detach from the Screw Shank. The recall affects 1,227 screws distributed nationwide.

    Product
    LineSider Spinal System 5.5/6.0mm, Non-mod Standard 7.1mm - Dual Lead, Non-cannulated. Thoracolumbosacral pedicle screw system; Spinal interlaminal fixation orthosis. Product sold as non-sterile and shipped in protective sterilization trays. Screws can also be shipped…
    Category
    Distribution
    16 states