FDA (Devices) recall action 91054

Primus Corporation dba Trinity Biotech recalled 2 products on 2022-12-07

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2022-12-07
Critical
0
Units
522

Why these products were recalled

The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. The unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.

The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. Additionally, per the firm, the unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • ModerateFDA (Devices)··2022-12-07

    Hemoglobin A1c Control Material Produces Incorrect Test Results

    A. MENARINI diagnostics hemoglobin A1c control materials (Lot 12161) are producing incorrect results that cause analyzer errors and prevent diagnostic testing. Facilities using this lot should discontinue use and contact the manufacturer.

    Product
    A. MENARINI diagnostics Hb 9210 PREMIER, LEVEL1, LEVEL II, Hemoglobin A1c Controls for hemoglobin A1c Assay, REF 45052, part number 05-01-0926, A1c analysis
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2022-12-07

    Medical diagnostic control kit recalled for out-of-range quality control results

    Primus Corporation recalls Trinity Biotech HbA1c Control Kits because Control Level I material produces out-of-range results that cause analyzer malfunction and testing delays, potentially delaying patient results.

    Product
    Trinity Biotech HbA1c (GHb) Contorls Kit, 500uL (Levels I & II), REF 01-04-0020, Part number 05-01-0040B, A1c analysis