FDA (Devices) recall action 91021

ARROW INTERNATIONAL Inc. recalled 31 products on 2022-12-21

The FDA (Devices) publishes a recall like this one as 31 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 31 pages.

Products
31
Announced
2022-12-21
Critical
1
Units
4,642

Why these products were recalled

There is a potential issue with short battery run-times on the affected intra-aortic balloon pump (IABP) devices.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–25 of 31

  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT2 Intra-Aortic Balloon Pump Battery Runtime Issue

    Arrow International is recalling the AutoCAT2 Intra-Aortic Balloon Pump due to a potential issue with short battery run-times. All lot and serial numbers worldwide are affected.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump, AC3 IABP NA/AJLA, REF IAP-0601(IPN916548), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT2 Intra-Aortic Balloon Pump Battery Run-Time Issue

    ARROW INTERNATIONAL is recalling Arrow AutoCAT2 Intra-Aortic Balloon Pumps due to a potential issue with short battery run-times. Affected devices distributed worldwide may have reduced battery endurance.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump, AC3 IABP NA/EMEA, REF IAP-0600 (IPN917287), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT2 Intra-Aortic Balloon Pump Recalled for Short Battery Run-Times

    Arrow AutoCAT2 intra-aortic balloon pumps are being recalled due to a potential issue with short battery run-times. The recall affects 12 units distributed worldwide.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump, AEROAUTOCAT2, REF IAP-0435 (IPN000312), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AC3 Optimus cardiac pump recalled due to battery issues

    Arrow AC3 Optimus Intra-Aortic Balloon Pumps worldwide are being recalled due to short battery run-times. All lot and serial numbers are affected.

    Product
    Arrow AC3 Optimus Intra-Aortic Balloon Pump, AC3 Optimus IABP NA/EMEA, REF IAP-0700(IPN917285), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT2 Cardiac Pump Recalled for Short Battery Run-Time Risk

    Arrow International is recalling the AutoCAT2 intra-aortic balloon pump (model AEROAUTOCAT 2 WAVE) due to a potential issue with short battery run-times that may affect patient cardiac support.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump, AEROAUTOCAT 2 WAVE, REF IAP-0535 (IPN000329), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT2WAVE Intra-Aortic Balloon Pump Recalled for Short Battery Run-Time

    Arrow AutoCAT2WAVE intra-aortic balloon pumps are being recalled due to a potential issue with short battery run-times. The Class I recall affects 48 units distributed worldwide.

    Product
    Arrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500F (IPN000323), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT2WAVE Intra-Aortic Balloon Pump recalled for short battery runtime

    Arrow AutoCAT2WAVE IABP devices have been recalled due to a potential issue with short battery run-times. The recall covers 112 units distributed worldwide.

    Product
    Arrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500E (IPN000322), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AC3 Optimus intra-aortic balloon pump short battery runtime

    Arrow AC3 Optimus cardiac pumps may experience short battery run-times. The FDA has issued a Class I recall for all units worldwide.

    Product
    Arrow AC3 Optimus Intra-Aortic Balloon Pump, AC3 Optimus REFURBISHED, REF IAP-0700X (IPN914274), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT2 Intra-Aortic Balloon Pump Recalled for Short Battery Run-Times

    Arrow AutoCAT2 intra-aortic balloon pumps may experience shorter battery run-times than specified, affecting device reliability in cardiac care. The FDA has issued a Class I recall for all affected units worldwide.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump, AC3 IABP NA/AJLA, REF IAP-0601 (IPN001109), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT2 cardiac pump recalled for short battery run-times

    Arrow International is recalling 2,678 Arrow AutoCAT2 intra-aortic balloon pumps worldwide due to a potential issue with short battery run-times. The FDA classified this as a Class I recall.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump, AUTOCAT2 WAVE, REF IAP-0500 (IPN000320), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT2WAVE intra-aortic balloon pump recalled for short battery run-times

    Arrow AutoCAT2WAVE intra-aortic balloon pumps are being recalled for a potential issue with short battery run-times. The FDA Class I recall affects 39 units distributed worldwide, manufactured by Arrow International Inc.

    Product
    Arrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500D (IPN000321), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AC3 Optimus Intra-Aortic Balloon Pump Battery Failure Recall

    Arrow International recalls AC3 Optimus intra-aortic balloon pumps worldwide due to short battery run-times. The FDA issued a Class I recall affecting all lot and serial numbers.

    Product
    Arrow AC3 Optimus Intra-Aortic Balloon Pump, AC3 Optimus IABP NA/AJLA, REF IAP-0701 (IPN917286), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT 2 Cardiac Pump Recalled for Potential Battery Runtime Issue

    Arrow International is recalling Arrow AutoCAT 2 intra-aortic balloon pumps worldwide due to a potential issue with battery run-times.

    Product
    Arrow AutoCAT 2 Intra-Aortic Balloon Pump, Arrow AutoCAT 2 IABP, REF IAP-0400F (IPN000306), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AC3 Optimus IABP Devices Recalled for Short Battery Run-Time Issue

    Arrow AC3 Optimus Intra-Aortic Balloon Pump devices are being recalled worldwide due to a potential issue with short battery run-times. The FDA issued a Class I recall for these devices.

    Product
    Arrow AC3 Optimus Intra-Aortic Balloon Pump, AC3 Optimus IABP NA/AJLA, REF IAP-0701 (IPN916549), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT 2 Cardiac Pump Recalled for Battery Runtime Issues

    Arrow AutoCAT 2 intra-aortic balloon pumps are being recalled due to a potential issue with short battery run-times. Affected users should contact the manufacturer for guidance.

    Product
    Arrow AutoCAT 2 Intra-Aortic Balloon Pump, Arrow AutoCAT 2 IABP, REF IAP-0400J (IPN000308), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    FDA Recalls Arrow AutoCAT2WAVE IABP for Short Battery Runtime

    The Arrow AutoCAT2WAVE intra-aortic balloon pump is being recalled worldwide due to a potential issue with short battery run-times. Healthcare facilities should contact the manufacturer for guidance.

    Product
    Arrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500I (IPN000324), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Intra-Aortic Balloon Pump Battery Run-Time Issue Recall

    Arrow AutoCAT2WAVE intra-aortic balloon pumps are recalled due to a potential issue with short battery run-times. Fifteen units distributed worldwide are affected.

    Product
    Arrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500SV (IPN000327), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT2WAVE Intra-Aortic Balloon Pump Recalled for Battery Runtime Issues

    Arrow International is recalling 127 units of the AutoCAT2WAVE Intra-Aortic Balloon Pump worldwide due to a potential issue with battery runtime.

    Product
    Arrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500J (IPN000325), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Intra-Aortic Balloon Pump with Potential Short Battery Run-Times

    Arrow AutoCAT2WAVE Intra-Aortic Balloon Pumps are being recalled worldwide due to short battery run-times affecting device operation. Patients using this device should contact their healthcare provider immediately.

    Product
    Arrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500EX (IPN000322), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT2 Intra-Aortic Balloon Pump Recalled for Short Battery Run-Times

    Arrow International is recalling the AutoCAT2 Intra-Aortic Balloon Pump due to a potential issue with short battery run-times that could affect the device's ability to provide continuous cardiac support.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump, AC3 IABP NA/AJLA, REF IAP-0601(IPN917288), cardiac pump
    Category
    Distribution
    0 states
  • CriticalFDA (Devices)··2022-12-21

    Arrow AC3 Optimus IABP recalled for shortened battery run-times

    Arrow AC3 Optimus Intra-Aortic Balloon Pump (IABP) devices are recalled worldwide due to a potential issue with shortened battery run-times. All lot and serial numbers of model IAP-0700 are affected.

    Product
    Arrow AC3 Optimus Intra-Aortic Balloon Pump, AC3 Optimus IABP NA/EMEA, REF IAP-0700 (IPN001112), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT2WAVE Intra-Aortic Balloon Pump Recalled for Short Battery Run-Times

    Arrow International is recalling Arrow AutoCAT2WAVE Intra-Aortic Balloon Pump devices worldwide due to a potential issue with short battery run-times.

    Product
    Arrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-05001 (IPN000333), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT2 Intra-Aortic Balloon Pump recalled for short battery run-times

    Arrow International is recalling 1,394 units of the Arrow AutoCAT2 Intra-Aortic Balloon Pump worldwide due to a potential issue with short battery run-times.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump, AUTOCAT 2 IAPB, REF IAP-0400 (IPN000302), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Intra-Aortic Balloon Pump Recalled for Potential Battery Run-Time Issue

    Arrow AutoCAT2WAVE Intra-Aortic Balloon Pump devices are being recalled worldwide due to a potential issue with short battery run-times. Healthcare providers and patients should contact the manufacturer for guidance.

    Product
    Arrow AutoCAT2WAVE Intra-Aortic Balloon Pump, Arrow AutoCAT2WAVE IABP, REF IAP-0500J (IPN000334), cardiac pump
    Category
    Distribution
    0 states
  • SevereFDA (Devices)··2022-12-21

    Arrow AutoCAT2 Intra-Aortic Balloon Pump Battery Runtime Recall

    The Arrow AutoCAT2 intra-aortic balloon pump has been recalled due to potential short battery run-times. This Class I recall affects devices distributed worldwide.

    Product
    Arrow AutoCAT2 Intra-Aortic Balloon Pump, AEROAUTOCAT2 WAVE REFURBISHED, REF IAP-0535X (IPN000329), cardiac pump
    Category
    Distribution
    0 states