FDA (Devices) recall action 90967

Carefusion 2200 Inc recalled 45 products on 2022-12-14

The FDA (Devices) publishes a recall like this one as 45 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 45 pages.

Products
45
Announced
2022-12-14
Critical
0
Units
12,625

Why these products were recalled

BD has determined that the Instructions For Use (IFU) provided by BD is missing certain content that is contained in the manufacturer's IFU. This missing content is related to the interface with the device power supply, as well as cleaning and maintenance instructions. Refer to the attached V. Mueller" Bipolar Forcep IFU.

XXX

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

26–45 of 45

  • ModerateFDA (Devices)··2022-12-14

    Carefusion Bipolar Surgical Forceps Recalled for Missing Instructions

    Carefusion 2200 Inc. is recalling 41 units of Hardy Bayonet Irrigating Bipolar Forceps due to incomplete instructions for use. The supplied instructions lack content about device power supply interface and maintenance procedures.

    Product
    Carefusion V. Mueller HARDY BAYONET IRRIGATING BIPOLAR FORCEPS INSULATED, 0.5MM TIP OVERALL LENGTH 8-1/2" (215MM), REF F-1335
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2022-12-14

    Bipolar Forceps Recall Due to Incomplete Instructions for Use

    Carefusion is recalling 259 surgical bipolar forceps units because the Instructions for Use are missing critical information about power supply interface and maintenance procedures.

    Product
    Carefusion V. Mueller GERALD BAYONET BIPOLAR FORCEPS 1.0MM TIP, INSULATED OVERALL LENGTH 7-1/2" (190MM), REF F-1113
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-12-14

    Carefusion Bipolar Forceps Recalled for Incomplete Instructions for Use

    BD is recalling Carefusion V. Mueller Bipolar Forceps due to incomplete Instructions for Use lacking critical content about power supply interface and cleaning procedures.

    Product
    Carefusion V. Mueller CUSHING BIPOLAR FORCEPS 1.5MM TIP, INSULATED OVERALL LENGTH 7" (180MM), REF F-2002
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-12-14

    Bipolar Forceps Recalled for Incomplete Instructions for Use

    Carefusion bipolar surgical forceps recalled due to incomplete Instructions for Use missing critical content on power supply interface and maintenance procedures. Distributed nationwide and internationally.

    Product
    Carefusion V. Mueller GERALD BIPOLAR FORCEPS 1.0MM WIDE TIP OVERALL LENGTH 7" (17.8CM) , REF F-1110
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-12-14

    Bipolar Forceps Recalled for Missing Instructions on Cleaning and Power Supply

    CareFusion V. Mueller bipolar forceps (Ref F-5073) recalled due to missing critical instructions for power supply interface and cleaning and maintenance procedures. Users should obtain the complete Instructions For Use from the manufacturer.

    Product
    CareFusion V. Mueller ROUND HANDLE BAYONET BIPOLAR FORCEPS 1.5MM TIP OVERALL LENGTH 10-1/2" (265MM), REF F-5073
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2022-12-14

    Bipolar Forceps Recall: Missing Instructions for Power Supply and Maintenance

    CareFusion bipolar forceps are being recalled due to missing Instructions for Use content about power supply interface, cleaning, and maintenance. Users should obtain the complete manufacturer instructions immediately.

    Product
    CareFusion V. Mueller ROUND HANDLE BAYONET BIPOLAR FORCEPS, 1.0MM TIP OVERALL LENGTH 10-1/2" (265MM), REF F-5069
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-12-14

    Surgical Bipolar Forceps Recall: Incomplete Instructions for Use

    Carefusion is recalling 218 units of surgical bipolar forceps due to incomplete product instructions. The Instructions For Use is missing information about power supply interface and cleaning/maintenance procedures.

    Product
    Carefusion V. Mueller HARDY BAYONET BIPOLAR FORCEPS WITH STOP, 0.5MM TIP OVERALL LENGTH 8-1/2" (215MM), REF F-1030
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-12-14

    Carefusion Bipolar Forceps Recalled Due to Incomplete Instructions

    Carefusion recalled 84 units of HARDY-STYLE BAYONET BIPOLAR FORCEPS nationwide and internationally due to incomplete Instructions for Use. The distributed documentation is missing information about power supply interface, cleaning, and maintenance procedures.

    Product
    Carefusion V. Mueller" HARDY-STYLE BAYONET BIPOLAR FORCEPS, 0.5MM TIP, INSULATED OVERALL LENGTH 8-1/2" (215MM) , REF F-1035
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2022-12-14

    Bipolar Irrigating Forceps Recall: Missing Power Supply and Maintenance Instructions

    Carefusion is recalling 132 units of V. Mueller bipolar irrigation forceps due to incomplete instruction manuals missing power supply interface and maintenance guidance.

    Product
    Carefusion V. Mueller SCOVILLE-GREENWOOD BAYONET IRRIGATING BIPOLAR FORCEPS INSULATED, 1.5MM TIP OVERALL LENGTH 7-3/4" (195MM), REF F-1305
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-12-14

    CareFusion Bipolar Forceps Recalled Due to Missing Instructions for Use

    CareFusion V. Mueller bipolar forceps are being recalled because the provided Instructions for Use are missing critical content about power supply interface and cleaning/maintenance. The complete information is available in the manufacturer's full IFU.

    Product
    CareFusion V. Mueller TITANIUM BAYONET BIPOLAR FORCEPS, 1.0MM TIP, STRAIGHT MEDIUM OVERALL LENGTH 8-3/4" (225MM), REF F-5126
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-12-14

    Bipolar forceps recall for incomplete operating and maintenance instructions

    Carefusion is recalling neurosurgical bipolar forceps due to incomplete Instructions for Use. The product is missing critical information about power supply interface and cleaning/maintenance procedures.

    Product
    Carefusion V. Mueller Neuro/Spine JEWELERS BIPOLAR FORCEPS 0.4MM TIP, CURVED, # 7 OVERALL LENGTH 4-3/4" (120MM), REF F-3007
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-12-14

    Surgical Bipolar Forceps Recalled for Missing Safety Instructions

    Carefusion is recalling surgical bipolar forceps due to incomplete instructions for use. The missing information covers power supply interface and cleaning and maintenance procedures.

    Product
    Carefusion V. Mueller ADSON BIPOLAR FORCEPS 1.0MM TIP OVERALL LENGTH 4-3/4" (120MM), REF F-3010
    Category
    Distribution
    0 states
  • LowFDA (Devices)··2022-12-14

    Medical Device Recall: Missing Instructions for Bipolar Surgical Forceps

    CareFusion bipolar forceps recalled because the provided instructions are missing critical information about power supply interface and cleaning procedures. No injuries reported.

    Product
    CareFusion V. Mueller TITANIUM ROUND HANDLE BAYONET BIPOLAR FORCEPS,1.0MM TIP OVERALL LENGTH 10-1/2" (265MM), REF F-5018
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-12-14

    Bipolar Forceps Recalled for Missing Power Supply and Cleaning Instructions

    Carefusion 2200 Inc is recalling 203 units of V. Mueller SEMKIN Bipolar Forceps due to incomplete Instructions for Use. The IFU is missing critical content related to power supply interface and cleaning/maintenance procedures.

    Product
    Carefusion V. Mueller" SEMKIN BIPOLAR FORCEPS WITH STOP, 0.7MM TIP OVERALL LENGTH 5-1/2" (140MM), REF F-1040
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-12-14

    Bipolar Surgical Forceps Recalled for Incomplete Instructions for Use

    Carefusion is recalling 418 units of bipolar surgical forceps because the provided instructions lack critical content about power supply interface and maintenance. Complete manufacturer instructions are necessary for safe device operation.

    Product
    CareFusion V. Mueller SCOVILLE-GREENWOOD BAYONET BIPOLAR FORCEPS 1.5MM TIP OVERALL LENGTH 7-3/4" (195MM) , REF F-1002
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2022-12-14

    CareFusion Bipolar Surgical Forceps Recalled for Incomplete Instructions

    CareFusion bipolar surgical forceps are being recalled because their instruction manual is missing critical information about power supply interface and cleaning procedures. BD is providing corrected instructions to healthcare facilities.

    Product
    CareFusion V. Mueller TITANIUM BAYONET IRRIGATING BIPOLAR FORCEPS INSULATED STRAIGHT FINE, 0.5MM TIP OVERALL LENGTH 8-3/4" (225MM), REF F-5302
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2022-12-14

    Bipolar Forceps Device Incomplete Instructions for Use Recalled

    CareFusion bipolar forceps (REF F-5061) recalled due to incomplete Instructions for Use missing device power supply interface and cleaning instruction content.

    Product
    CareFusion V. Mueller ROUND HANDLE BAYONET BIPOLAR FORCEPS 1.0MM TIP OVERALL LENGTH 9-1/2" (240MM), REF F-5061
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-12-14

    Bipolar Forceps Recall: Missing Device Use and Maintenance Instructions

    CareFusion bipolar forceps recalled due to incomplete Instructions for Use. Missing documentation covers power supply interface and cleaning/maintenance procedures.

    Product
    CareFusion V. Mueller TITANIUM ROUND HANDLE BAYONET BIPOLAR FORCEPS 0.7MM TIP OVERALL LENGTH 10-1/2" (265MM), REF F-5016
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2022-12-14

    Bipolar Surgical Forceps Recalled for Incomplete Instructions for Use

    Carefusion bipolar surgical forceps recalled nationwide due to incomplete Instructions for Use. The distributed IFU is missing guidance on power supply interface and cleaning and maintenance procedures.

    Product
    Carefusion V. Mueller GERALD BAYONET BIPOLAR FORCEPS 1.0MM TIP, INSULATED IRRIGATING OVERALL LENGTH 7-1/2" (190MM), REF F-1303
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-12-14

    Bipolar Forceps Recall: Incomplete Instructions for Surgical Device

    BD recalls CareFusion bipolar forceps (Ref F-5301) due to incomplete Instructions For Use. The IFU is missing information about power supply interface and cleaning and maintenance.

    Product
    CareFusion V. Mueller TITANIUM BAYONET INSULATED IRRIGATING BIPOLAR FORCEPS 1.5MM TIP, STRAIGHT, INSULATED OVERALL LENGTH 8-3/4" (225MM), REF F-5301
    Category
    Distribution
    0 states