Why these products were recalled

The subject products are being recalled because they contain the incorrect size drill guide and will not prepare the bone correctly for the size of the implant included in the sterile kit. If the incorrect drill guide is used, there is potential for bone damage due to incorrectly placed drill holes for the implant included in the kit.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2022-10-26

    Surgical Implant Kit Recall: Incorrect Drill Guide Included

    Synthes surgical implant kits are being recalled because they contain an incorrect size drill guide that cannot properly prepare the bone for the implant, creating a risk of bone damage during surgery.

    Product
    SPEED 15X12X12mm Continuous Compression Implant, Product No. SE-1512
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-26

    Synthes SPEED implant recall: drill guide size mismatch in surgical kit

    Synthes is recalling SPEED 9X10X10mm surgical implant kits nationwide due to an incorrect-size drill guide that will not properly prepare bone for the implant, creating potential for bone damage.

    Product
    SPEED 9X10X10mm Continuous Compression Implant, Product No. SE-0910
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-26

    DePuy Synthes Hammertoe Implants Recalled for Incorrect Drill Guide Size

    Synthes (USA) Products LLC is recalling 196 DePuy Synthes Hammertoe Continuous Compression Implants due to an incorrect drill guide size that will not properly prepare bone for the implant, creating potential for bone damage.

    Product
    DePuy Synthes Hammertoe Continuous Compression Implants, Product No. 46.239.001
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-26

    Synthes Orthopedic Implant Drill Kit Recall: Incorrect Drill Guide Size

    Synthes (USA) Products LLC is recalling 81 SPEEDSHIFT20x20x20 Offset 10mm Implant Drill Kits nationwide due to incorrect drill guide sizing. The defect could result in bone damage and incorrectly placed drill holes during orthopedic surgery.

    Product
    SPEEDSHIFT20x20x20 Offset 10mm Implant Drill Kit, Product No. SE-2020-10. Used with DePuy Synthes SPEEDSHIFT Continuous Compression Staple Implants for bone fixation and management of fracture and reconstructive surgery.
    Category
    Distribution
    0 states