FDA (Devices) recall action 90867

SIEMENS HEALTHCARE DIAGNOSTICS INC recalled 2 products on 2022-10-26

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2022-10-26
Critical
0
Units
10,574

Why these products were recalled

N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2022-10-26

    Siemens N Antiserum to Human IgG Recall for High-Dose Performance Failure

    Siemens Healthcare Diagnostics recalls N Antiserum to Human IgG reagent for laboratory analyzers because it does not meet expected High-Dose Hook Effect performance specifications for IgG measurements in certain sample types.

    Product
    N Antiserum to Human IgG, 2mL Variant for BN II System and BN ProSpec System-IVD Siemens Material Number (SMN): 10446297
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-26

    N Antiserum to Human IgG Fails High-Dose Hook Effect Performance Standards

    Siemens N Antiserum to Human IgG reagent batches are not meeting performance standards for high-dose immunoassay testing. The issue affects 10,201 packs distributed nationwide and may result in inaccurate diagnostic test results.

    Product
    N Antiserum to Human IgG, 5mL Variant for BN II System and BN ProSpec System- IVD Siemens Material Number (SMN): 10446299
    Category
    Distribution
    0 states