The Recall Desk

FDA (Devices) recall action 90818

Leica Microsystems, Inc. recalled 4 products on 2022-10-05

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2022-10-05
Critical
0
Units
202

Why these products were recalled

During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–4 of 4

  • HighFDA (Devices)··2022-10-05

    Leica Surgical Operating Microscopes ARveo 8 recalled for illumination accuracy issue

    Leica Microsystems is recalling 30 ARveo 8 surgical microscopes (Part Number 10449063) distributed nationwide due to a photodiode component change causing inaccurate illumination adjustment when using the BrightCare Plus Luxmeter feature.

    Product
    Leica Surgical Operating Microscopes, ARveo 8, Part Number 10449063
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Leica M530 Surgical Microscopes recalled for illumination adjustment defect

    Leica Microsystems is recalling 115 M530 surgical microscopes due to a photodiode component change that causes inaccurate illumination limit adjustment when using the BrightCare Plus feature with Luxmeter.

    Product
    Leica Surgical Operating Microscopes, M530 OHX, Part Number 10448737
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Leica Surgical Operating Microscopes recalled for illumination adjustment defect

    Leica Microsystems recalls 25 Leica Surgical Operating Microscopes (Part Number 10448999) due to a photodiode component change causing inaccurate illumination adjustment when BrightCare Plus with Luxmeter is used.

    Product
    Leica Surgical Operating Microscopes, Arveo, Part Number 10448999
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2022-10-05

    Leica Surgical Microscopes: Photodiode Defect Causes Inaccurate Illumination

    Leica Microsystems is recalling 32 surgical operating microscopes due to a photodiode component defect that produces inaccurate illumination adjustment when BrightCare Plus with Luxmeter is used.

    Product
    Leica Surgical Operating Microscopes, PROvido Part Numbers: a. 10448976, b. 10448950
    Category
    Distribution
    0 states