The Recall Desk

FDA (Devices) recall action 90793

Cook Incorporated recalled 24 products on 2022-10-05

The FDA (Devices) publishes a recall like this one as 24 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 24 pages.

Products
24
Announced
2022-10-05
Critical
0
Units
1,814

Why these products were recalled

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–24 of 24

  • HighFDA (Devices)··2022-10-05

    Cope Mandril Wire Guide Guidewires Recalled for Packaging Seal Breach

    Cook Incorporated is recalling Cope Mandril Wire Guide stainless steel guidewires because the packaging chevron seal may be completely breached, compromising device sterility. Affected lots total 2,748 units in the US and 302 internationally.

    Product
    Cope Mandril Wire Guide (Stainless Steel), Stainless Steel Guidewire Reference Part Number/GPN PMG-18SP-60-COPE G05183 PMG-18SP-100-COPE G05822 PMG-18SP-125-COPE G10077
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Cook Roadrunner Guidewire Recalled for Compromised Sterile Packaging Seal

    Cook Incorporated is recalling Roadrunner PC Wire Guide guidewires due to a complete breach of the sterile packaging seal. The breach may compromise device sterility, affecting 202 units in the US and 11 internationally.

    Product
    Roadrunner PC Wire Guide (Nimble Floppy), Guidewire RFPC-35-145 G07914
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Cook Roadrunner PC Guidewire Recalled for Compromised Packaging Seal

    Cook Incorporated recalls Roadrunner PC Wire Guide devices due to a potential packaging seal breach that may compromise sterility. The recall affects 115 units distributed domestically and internationally.

    Product
    Roadrunner PC Wire Guide (The Firm LT), Guidewire Reference Part Number/GPN RLPC-35-145 G06866
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Bentson Wire Guide Devices Recalled for Potential Sterility Compromise

    Cook Incorporated is recalling Bentson Wire Guide devices due to potential packaging defects that may compromise device sterility.

    Product
    Bentson Wire Guide, Guidewire (OUS ONLY); Reference Part Number: TSMB-35-145 (GPN G27041)
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Cook Medical Wire Guide Devices Recalled for Packaging Breach

    Cook Incorporated is recalling specific lots of Double Flexible Tipped Wire Guide Guidewire devices due to a complete breach of the packaging chevron seal, which may compromise sterility.

    Product
    Double Flexible Tipped Wire Guide, Guidewire Reference Part Number/GPN TDOC-35-145-0-3 G00362 C-TDOC-18-50-0-2 G05146 TDOC-35-50-0-3 G00364 C-DOC-21-40-0-2 G02431 C-TDOC-21-50-0-2 G05195 C-TDOC-25-50-0-2 G05145
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Cook Roadrunner PC Guidewires Recalled for Compromised Packaging Sterility

    Cook Incorporated is recalling Roadrunner PC Hydrophilic Guidewires due to potential packaging seal breaches that may compromise device sterility. Affected devices were distributed in the United States and internationally.

    Product
    Roadrunner PC Hydrophilic Wire Guide, Hydrophilic Guidewire Reference Part Number/GPN: RPC-035145-0-5 G34131 RPC-035145 G18154 RPC-035145-5 G34132 RPC-038145-0-5 G34129 RFSPC-038145-0 G17539 RFSPC-035145 G17540 RPC-038145-0 G18155 RPC-035145-0 G18153 RFSPC-035145-0…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Coons Interventional Guidewire Recall Due to Packaging Sterility Breach

    Cook Incorporated is recalling Coons Interventional Wire Guide devices because the chevron seal on packaging may be completely breached, compromising sterility. 1,236 units were distributed.

    Product
    Coons Interventional Wire Guide, Guidewire; Reference Part Number/GPN: THSF-35-145-COONS G02356 THSF-38-145-COONS G02324 THSF-35-180-COONS G02621 THSF-35-80-COONS G12268
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    FDA Recalls Cook Roadrunner PC Wire Guides for Packaging Sterility Breach

    Cook Incorporated is recalling Roadrunner PC Wire Guide devices due to a breach in packaging's chevron seal that may compromise sterility. Affected devices should not be used.

    Product
    Roadrunner PC Wire Guide (Nimble), Guidewire Reference Part Number/GPN: RFSPC-35-145 G09607
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Roadrunner PC Wire Guide Recalled for Packaging Breach and Sterility Compromise

    Cook Incorporated recalls Roadrunner PC Wire Guide devices due to packaging seal breaches that may compromise sterility. Devices were distributed nationwide and internationally.

    Product
    Roadrunner PC Wire Guide (The Firm), Guidewire Reference Part Number/GPN RPC-35-145 G06979 RPC-35-80 G09534
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Cope Nitinol Guidewire Recalled Due to Packaging Sterility Breach

    Cook Incorporated is recalling Cope Mandril Wire Guide (Nitinol) devices because the packaging's chevron seal may be completely breached, potentially compromising sterility. The recall affects 1,501 US units and 380 international units.

    Product
    Cope Mandril Wire Guide (Nitinol), Nitinol Guidewire; Reference Part Number/GPN: PMG-18SP-60-COPE-NT-ST G09395 PMG-18SP-100-COPE-NT G08687 PMG-18SP-60-COPE-NT G08427
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Bentson Wire Guide Recalled for Packaging Defect Affecting Sterility

    Cook Incorporated is recalling Bentson Wire Guide guidewires due to packaging seal failures that may compromise sterility. The recall affects over 5,500 units distributed in the U.S. and internationally.

    Product
    Bentson Wire Guide (20cm Distal Flexibility with Softened Tip), Guidewire; Reference Part Number: TSFB-35-80 (GPN G01315) TSFB-35-145 (GPN G00691) TSFB-35-200 (GPN G01095)
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Bentson Plus Guidewire Recall for Compromised Packaging Sterility Seal

    Cook Incorporated is recalling Bentson Plus Wire Guides due to potential packaging seal breaches that may compromise device sterility. Affected devices may not maintain proper sterility before use.

    Product
    Bentson Plus Wire Guide (Extra Body/20cm Distal Flexibility with Softened Tip), Guidewire; Reference Part Number: TSFBP-35-180 (GPN G09782) TSFBP-35-145 (GPN G09735)
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Cook Movable Core Wire Guide Guidewires Recall Due to Packaging Defect

    Cook Incorporated is recalling specific lots of Movable Core Wire Guide guidewires due to a potential breach of the packaging seal that may compromise device sterility.

    Product
    Movable Core Wire Guide (Tapered Tefcor) (OUS ONLY), Guidewire Reference Part Number/GPN TCMTNA-35-145-3 G01678
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Vinyl Connecting Tubes Recalled Due to Packaging Seal Breach

    Cook Incorporated is recalling Vinyl Connecting Tubes because of a complete breach of the chevron seal in the packaging, which may compromise device sterility.

    Product
    Vinyl Connecting Tube, Connecting Tube Reference Part Number/GPN CTU14.0-30-ST G02278 CTU14.0-40-ST G02327 CTU14.0-30 G02791 CTU14.0-50-ST G02464 C-CTU14.0-30-ST G02898
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Cook Incorporated Recalls Guidewire Due to Packaging Seal Breach

    Cook Incorporated is recalling Heavy Double Flexible Tipped Wire Guide devices due to potential packaging seal breach that may compromise sterility. Affected products were distributed domestically and worldwide.

    Product
    Heavy Double Flexible Tipped Wire Guide, Guidewire Reference Part Number/GPN: THDOC-35-60-0-3 G11949 C-HDOC-18-40-0-2 G09201 C-THDOC-18-40-0-2-HTS G26925
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Cook Medical Guidewire Recall: Packaging Seal Breach May Compromise Sterility

    Cook Incorporated is recalling Fixed Core Wire Guide devices due to a potential breach in the packaging seal that may compromise device sterility.

    Product
    Fixed Core Wire Guide (Safe-T-J Curved), Guidewire Reference Part Number/GPN: TSCF-25-180-3 G01602 TSCF-35-180-3 G00902 TSCF-21-260-3 G00455 THSCF-25-260-3 G02384 TSCF-25-145-3 G00466 THSCF-35-80-3 G04069 THSCF-35-180-15 G01151 TSCF-21-50-3 G00456 THSCF-35-145-3 G00412…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Cook Guidewire Products Recalled Due to Compromised Sterile Packaging

    Cook Incorporated is recalling Fixed Core Wire Guide products due to potential breach of sterile packaging seals. Affected devices may have compromised sterility.

    Product
    Fixed Core Wire Guide, Guidewire Reference Part Number/GPN THSCF-25-180-3 G02165 TSCF-25-80-3 G00476
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Cook Lunderquist Ring Torque Wire Guide packaging breach may compromise sterility

    Cook Incorporated is recalling Lunderquist Ring Torque Wire Guides due to a complete breach in the chevron seal of the packaging, which may compromise device sterility. The recall affects multiple lot numbers distributed nationwide and internationally.

    Product
    Lunderquist -Ring Torque Wire Guide, Guidewire Reference Part Number/GPN THSF-35-125-THG G27031 THSF-35-145-THG G27033
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Cook Newton Wire Guide Recalls Due to Potential Sterility Compromise

    Cook Incorporated is recalling certain Newton Wire Guide surgical guidewires because packaging seals may be breached, potentially compromising device sterility.

    Product
    Newton Wire Guide (LT Taper/10cm Distal Flexibility), Guidewire Reference Part Number/GPN: TSFNA-35-145 G00701 TSCFNA-35-145-3 G00561
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Torq-Flex Wire Guide Guidewires Recalled Due to Packaging Seal Defect

    Cook Incorporated is recalling Torq-Flex Wire Guide guidewires because the packaging chevron seal may be breached, potentially compromising device sterility. The affected units were distributed domestically and internationally.

    Product
    Torq-Flex Wire Guide Australian Modification, Guidewire Reference Part Number/GPN STF-18-40-AUST G07304
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Medical Guidewire Recalled for Packaging Defect Affecting Sterility

    Cook Incorporated is recalling Movable Core Wire Guide (Tefcor) guidewires due to a packaging defect where the chevron seal may be breached, potentially compromising device sterility.

    Product
    Movable Core Wire Guide (Tefcor), Guidewire (OUS ONLY) Reference Part Number/GPN TMT-35-145 G01677
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Cook guidewires recalled for compromised sterility due to packaging breach

    Cook Incorporated is recalling Amplatz extra Stiff Wire Guides because packaging seals may be completely breached, potentially compromising device sterility.

    Product
    Amplatz extra Stiff Wire Guide, Stiff Guidewire (OUS ONLY); Reference Part Number: THSF-35-145-AES-SGH (GPN G06801) THSF-38-80-AES-SGH (GPN G06874)
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Cook Rosen Curved Wire Guide Recall: Sterility Concerns

    Cook Incorporated is recalling specific lots of Rosen Curved Wire Guide guidewires due to a complete breach of the packaging seal that may compromise device sterility. Affected devices were distributed nationwide in the US and worldwide.

    Product
    Rosen Curved Wire Guide, Guidewire Reference Part Number/GPN THSCF-35-220-1.5-ROSEN G01623 THSCF-35-80-1.5-ROSEN G01774 THSCF-35-180-1.5-ROSEN G01264 THSCF-35-145-1.5-ROSEN G01261
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-10-05

    Cook Guidewire Devices Recalled for Packaging Defect and Sterility Compromise

    Cook Incorporated is recalling fixed core wire guide devices due to packaging defects that may compromise device sterility. Approximately 5,669 units were distributed in the US and 5,286 internationally.

    Product
    Fixed Core Wire Guide (Straight), Guidewire Reference Part Number/GPN: C-SF-18-30 G02261 C-SF-25-40 G01937 TSF-18-260 G00590 C-SF-25-50 G01938 TSF-35-180 G00652 C-SF-18-40 G02255 TSF-25-145 G00617 TSF-32-145 G00638 C-SF-15-50 G02274 TSF-18-145 G00587 TSF-18-50 G00593…
    Category
    Distribution
    0 states