The Recall Desk

FDA (Devices) recall action 90771

Siemens Medical Solutions USA, Inc recalled 3 products on 2022-09-07

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2022-09-07
Critical
0
Units
—

Why these products were recalled

Siemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to release X-ray any longer until system shutdown.

The agency’s own wording, quoted from the enforcement report.

1–3 of 3

  • HighFDA (Devices)··2022-09-07

    Siemens Artis icono X-Ray Systems Recalled for Potential Hardware Failure

    Siemens has recalled Artis icono X-ray imaging systems due to a potential hardware issue affecting a specific lot of X-ray tubes. If the error detection mechanism fails, the system cannot release X-rays until it is shut down.

    Product
    Artis icono biplane- Model No. 11327600 Artis icono floor- Model No. 11327700
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-09-07

    X-Ray Imaging System X-Ray Tube Hardware Defect Worldwide Recall

    Siemens has recalled Artis Q and Artis Q.zen X-ray imaging systems due to a potential X-ray tube hardware defect. In rare cases, if error detection fails, the system will not be able to release X-ray until shut down; 1,389 units are affected globally.

    Product
    Artis Q floor- Model No. 10848280 Artis Q ceiling- Model No. 10848281 Artis Q biplane- Model No. 10848282 Artis Q zeego- Model No. 10848283 Artis Q.zen floor- Model No. 10848353 Artis Q.zen ceiling- Model No. 10848354 Artis Q.zen biplane- Model No. 10848355
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2022-09-07

    X-ray imaging system hardware defect may prevent operation

    Siemens is recalling certain X-ray imaging systems due to a potential hardware issue with the error detection mechanism. In rare cases, this could prevent X-ray emission until system shutdown.

    Product
    Artis pheno- Model No. 10849000
    Category
    Distribution
    0 states