FDA (Devices) recall action 90751

FUJIFILM Healthcare Americas Corporation recalled 3 products on 2022-09-21

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2022-09-21
Critical
0
Units
82

Why these products were recalled

Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the work list.result, may result in the need for additional scanning to capture the images properly for reconstruction, thereby exposing the patient to additional radiation

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2022-09-21

    FUJIFILM Surpria Imaging Software May Fail During Scanning

    Surpria software versions V3.11 and V3.22 may encounter processing errors that prevent image reconstruction, potentially requiring additional scans and exposing patients to unnecessary radiation.

    Product
    Surpria: Software Version V3.11, V3.22
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-09-21

    Surpria 64 Software Error May Prevent Image Reconstruction During Scanning

    Surpria 64 CT scanner software versions V3.11 and V3.22 may encounter errors that prevent image reconstruction, requiring repeat scans that expose patients to additional radiation. 16 units nationwide are affected.

    Product
    Surpria 64: Software Version V3.11, V3.22
    Category
    Distribution
    0 states