FDA (Devices) recall action 90471

Randox Laboratories Ltd. recalled 2 products on 2022-07-27

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2022-07-27
Critical
0
Units
24

Why these products were recalled

A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in non-numerical results. This affects Siemens Atellica CH analysers only. Customer have reported an atypical calibration curve that results in the values around the low QC range to produce a non-numerical result. It appears to be related to the bottle values of the calibrator producing a curve shape that is too linear for the logitlog calculation to calculate. This issue can only be …

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • ModerateFDA (Devices)··2022-07-27

    Laboratory Reagent Recalled Due to Inaccurate Calibration Curves

    Randox Laboratories is recalling Soluble Transferrin Receptor reagent (TF10159) due to calibration curve defects causing inaccurate test results on Siemens Atellica CH analyzers.

    Product
    Soluble Transferrin Receptors (STFR), Catalog Number TF10159 (Research Use Only). IVD
    Category
    Distribution
    1 state
  • ModerateFDA (Devices)··2022-07-27

    Soluble Transferrin Receptor Calibrator Recall Due to Lab Errors

    Randox Laboratories is recalling Soluble Transferrin Receptors Calibrator kits used in lab analyzers. The calibrators can produce incorrect test results, with some values potentially underestimated by up to 30 percent.

    Product
    Soluble Transferrin Receptors Calibrator Series (STFR CAL), Catalog Number TF10161. IVD
    Category
    Distribution
    1 state