FDA (Devices) recall action 90087

Beckman Coulter Inc. recalled 6 products on 2022-05-25

The FDA (Devices) publishes a recall like this one as 6 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 6 pages.

Products
6
Announced
2022-05-25
Critical
0
Units
125

Why these products were recalled

WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–6 of 6

  • HighFDA (Devices)··2022-05-25

    Medical analyzer instruments recalled for incorrect power connector rating

    Beckman Coulter is recalling DxM 1096 MicroScan WalkAway Instruments because some units may have an incorrect AC power entry connector rated at 1 ampere instead of 10 amperes, posing a potential electrical safety risk.

    Product
    DxM 1096 MicroScan WalkAway Instrument REF B1018-496
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2022-05-25

    MicroScan WalkAway-40 Medical Instruments Recalled for Incorrect Power Connector Rating

    Three MicroScan WalkAway-40 medical instruments were recalled because they have an AC power connector rated at 1 ampere instead of the correct 10 ampere rating. This mismatch poses a potential electrical and fire hazard.

    Product
    MicroScan WalkAway-40 plus Reconditioned Instrument REF B1018-340R
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2022-05-25

    Medical diagnostic instrument recalled for electrical connector assembly defect

    Beckman Coulter is recalling 40 DxM 1040 MicroScan WalkAway laboratory instruments due to an assembly defect: some units have an AC power connector inlet rated at 1 ampere instead of the correct 10-ampere rating, which could impair power delivery.

    Product
    DxM 1040 MicroScan WalkAway Instrument REF B1018-440
    Category
    Distribution
    31 states
  • HighFDA (Devices)··2022-05-25

    MicroScan WalkAway-40 plus Instrument recalled for incorrect AC power connector

    Beckman Coulter is recalling 22 MicroScan WalkAway-40 plus Instruments with an AC power entry connector rated at 1 ampere instead of the correct 10 amperes, creating a potential electrical hazard.

    Product
    MicroScan WalkAway-40 plus Instrument REF B1018-283
    Category
    Distribution
    31 states