FDA (Devices) recall action 90076

GE Healthcare, LLC recalled 2 products on 2022-06-15

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2022-06-15
Critical
0
Units
123,319

Why these products were recalled

Insufficient battery backup power resulting in premature shutdown of the ventilator when not connected to AC mains power supply.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • SevereFDA (Devices)··2022-06-15

    Ventilator Backup Battery Failure Recalled by GE Healthcare Worldwide

    GE Healthcare is recalling backup batteries for ventilators used in critical care. The batteries may not provide sufficient backup power when disconnected from AC, potentially causing unexpected device shutdown.

    Product
    Replacement back up batteries distributed on or after April 1, 2019, for CARESCAPE R860, Engstr¿om Carestation and Engstr¿om PRO Ventilators.
    Category
    Distribution
    38 states
  • SevereFDA (Devices)··2022-06-15

    GE Healthcare Ventilators Recalled for Insufficient Battery Backup Power

    GE Healthcare is recalling multiple ventilator models due to insufficient battery backup power. The affected units may shut down prematurely when not connected to AC power.

    Product
    Back up batteries in Ventilators labeled as the following: a. CARESCAPE R860 Ventilators (K142679) b. ENGSTROM (K111116) c. ENGSTROM Carestation (K111116) d. ENGSTROM PRO (K111116)
    Category
    Distribution
    38 states