Siemens Medical Solutions USA, Inc recalled 4 products on 2022-05-04

The FDA (Devices) publishes a recall like this one as 4 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 4 pages.

Products
4
Announced
2022-05-04
Critical
0
Units
31

Why these products were recalled

Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–4 of 4

  • HighFDA (Devices)··2022-05-04

    CT Scanner Software Update May Display Incorrect Laser Offset Coordinates

    Siemens CT scanner software upgrade may cause incorrect laser offset coordinates to display during patient positioning. No injuries reported.

    Product
    syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Up Model Number 11061620. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.
    Category
    Distribution
    20 states
  • HighFDA (Devices)··2022-05-04

    CT scanner software upgrade causes incorrect laser position display

    Siemens CT scanner software upgrade may display incorrect laser offset coordinates during patient marking. This defect affects 15 systems in the US across multiple states.

    Product
    syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Sim Model Number 11061660. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.
    Category
    Distribution
    20 states
  • HighFDA (Devices)··2022-05-04

    CT Scanner Software Update May Display Incorrect Laser Positioning Coordinates

    Siemens CT scanner imaging software upgraded to version VB60 may display incorrect laser offset coordinates during patient marking. Fourteen units are affected in the United States.

    Product
    syngo.via RT Image Suite upgraded to VB60 running on SOMATOM go.Open Pro Model Number 11061670. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.
    Category
    Distribution
    20 states
  • HighFDA (Devices)··2022-05-04

    Radiation therapy software update displays incorrect laser positioning

    Siemens syngo.via RT imaging software upgraded to VB60 automatically modifies laser configuration, potentially displaying incorrect laser offset coordinates during radiation therapy patient marking.

    Product
    syngo.via RT Image Suite upgraded to VB60 running on syngo.via Model Number 10496180. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.
    Category
    Distribution
    20 states