The Recall Desk

FDA (Devices) recall action 89870

Siemens Medical Solutions USA, Inc recalled 3 products on 2022-04-27

The FDA (Devices) publishes a recall like this one as 3 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 3 pages.

Products
3
Announced
2022-04-27
Critical
0
Units
149

Why these products were recalled

Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential software errors are related to the block movements function and manual rotation of the flat detector. Depending on the status of the intervention, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where clinical treatment may be delayed until an additional restart is performed

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2022-04-27

    Medical imaging system software errors may delay treatment procedures

    Siemens Artis icono biplane diagnostic imaging systems may experience software errors affecting block movements and detector rotation, potentially delaying treatment until the system is restarted.

    Product
    Artis icono biplane- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single and biplane diagnostic imaging and interventional procedures. Model: 11327600
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-04-27

    Siemens Artis pheno diagnostic imaging system software malfunction recall

    Siemens Medical Solutions is recalling 20 Artis pheno diagnostic imaging systems nationwide due to software errors in block movement and detector rotation functions. The malfunction may limit device functionality and delay clinical treatment until the system restarts.

    Product
    Artis pheno- in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use-single and biplane diagnostic imaging and interventional procedures. Model: 10849000
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-04-27

    Diagnostic imaging system software malfunction may interrupt clinical procedures

    Siemens is recalling 35 Artis icono floor diagnostic imaging systems due to five potential software errors affecting equipment movements and detector control. The defects may limit functionality during procedures and could delay clinical treatment until the system is restarted.

    Product
    Artis icono floor in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single and biplane diagnostic imaging and interventional procedures. Model: 11327700
    Category
    Distribution
    0 states