Siemens Medical Solutions USA, Inc recalled 14 products on 2022-04-27

The FDA (Devices) publishes a recall like this one as 14 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 14 pages.

Products
14
Announced
2022-04-27
Critical
0
Units
2,866

Why these products were recalled

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–14 of 14

  • HighFDA (Devices)··2022-04-27

    X-ray Imaging System May Fail to Cool Tube Properly

    Siemens Artis zee biplane MN X-ray systems may have inadequate cooling in the X-ray tube circuit, causing automatic system shutdown during operation. Seven affected systems are distributed nationwide.

    Product
    Artis zee biplane MN, Model Number 10094143
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-04-27

    Artis zeego X-ray imaging system recalled for insufficient cooling

    Siemens is recalling 183 Artis zeego X-ray imaging systems due to a cooling system defect that may prevent adequate X-ray tube cooling, triggering automatic system shutdown.

    Product
    Artis zeego, Model Number 10280959
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-04-27

    X-ray Imaging System Coolant Deficiency May Disable Equipment Operation

    The Artis Q X-ray system may fail to adequately cool the X-ray tube if coolant levels drop, causing the system to disable imaging and display error messages. 418 systems are affected nationwide.

    Product
    Artis Q ceiling, Model Number 10848281
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-04-27

    Medical device X-ray imaging system recalled for cooling failure

    Siemens Medical Solutions is recalling its Artis Q.zen X-ray imaging system due to a coolant circulation defect that may cause insufficient tube cooling and system shutdown.

    Product
    Artis Q.zen floor, Model Number 10848353
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2022-04-27

    X-Ray Imaging System Cooling Circuit Failure Recall

    Siemens is recalling Artis zee floor MN X-ray systems due to a potential cooling circuit malfunction. If coolant drops below adequate levels, the X-ray tube may not cool properly and the system will automatically shut down.

    Product
    Artis zee floor MN, Model Number 10094142
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-04-27

    Artis Q floor X-ray system cooling failure poses overheating risk

    Siemens is recalling 80 Artis Q floor X-ray systems nationwide due to a cooling circuit failure that may cause insufficient cooling of the X-ray tube, resulting in automatic system shutdown.

    Product
    Artis Q floor, Model Number 10848280
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-04-27

    Medical X-ray imaging system recalled for cooling circuit malfunction

    Siemens' Artis zee X-ray imaging systems may have inadequate cooling if the coolant level drops, potentially causing X-ray tube overheating. The system displays warnings and blocks X-ray operation to prevent equipment damage.

    Product
    Artis zee multi-purpose, Model Number 10094139
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-04-27

    Artis Q zeego X-ray system recalled for inadequate cooling risk

    Siemens is recalling 38 Artis Q zeego X-ray systems nationwide due to a cooling circuit malfunction that may prevent adequate cooling of the X-ray tube, triggering automatic system shutdown.

    Product
    Artis Q zeego, Model Number 10848283
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-04-27

    Artis zee floor X-ray systems recalled for cooling system failure

    Siemens is recalling 305 Artis zee floor X-ray imaging systems because the cooling circuit may fail to maintain adequate coolant levels, potentially causing the X-ray tube to overheat. The system includes automatic safeguards that shut down X-ray operation if overheating is detected.

    Product
    Artis zee floor, Model Number 10094135
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2022-04-27

    X-ray Imaging System Cooling Failure May Cause Tube Overheating

    Siemens Artis zee ceiling X-ray systems may fail to cool the X-ray tube if the coolant level drops. The system will automatically block X-ray emissions and display warning messages to prevent equipment damage.

    Product
    Artis zee ceiling, Model Number 10094137
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-04-27

    Siemens Artis Q.zen ceiling system recall: inadequate X-ray tube cooling risk

    Siemens recalls 81 Artis Q.zen ceiling imaging systems distributed nationwide due to risk of inadequate X-ray tube cooling if coolant levels fall below a critical threshold, potentially triggering system shutdown.

    Product
    Artis Q.zen ceiling, Model Number 10848354
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-04-27

    Siemens Artis zee biplane X-ray system cooling defect recall

    Siemens is recalling 373 Artis zee biplane X-ray diagnostic systems due to a cooling circuit defect that could prevent adequate tube cooling. When the coolant level drops, the system may overheat and shut down imaging capabilities.

    Product
    Artis zee biplane, Model Number 10094141
    Category
    Distribution
    0 states
  • ModerateFDA (Devices)··2022-04-27

    Artis Q.zen Biplane X-ray Systems Coolant Level Failure Recall

    Siemens is recalling 69 Artis Q.zen biplane systems where coolant levels may drop sufficiently to prevent adequate X-ray tube cooling, triggering automatic shutdown safeguards.

    Product
    Artis Q.zen biplane, Model Number 10848355
    Category
    Distribution
    0 states