This correction is due to software anomalies occurring during use. If the operator initiates therapy by entering a value other than the default value for the patient Gain/Loss Limit or the Return Pressure Drop Limit (RDL) and uses the Same Patient function, the PrisMax System will return the patient Gain/Loss Limit or the RDL to the default value, rather than the non-default value entered at the initiation of treatment. The PrisMax System interface will display the original values once treatment is restarted but does not have an alert to inform the user of the change. The RDL value will be ret…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–3 of 3

  • HighFDA (Devices)··2022-05-04

    PrisMax System software anomaly affects therapy parameter settings

    The PrisMax System has a software defect that can silently revert custom therapy parameters to defaults when using the Same Patient function without notifying the operator.

    Product
    PrisMax System, Product Codes: a) 955558, b) 955627, c) 955701
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-05-04

    PrisMax System software causes unalerted treatment parameter reversion

    The PrisMax System may revert custom treatment parameters to default values without notifying the operator when the Same Patient function is used. This could cause patients to receive unintended treatment settings.

    Product
    PrisMax System, Product Code 955724
    Category
    Distribution
    0 states