St. Jude Medical, Cardiac Rhythm Management Division recalled 2 products on 2022-04-27

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2022-04-27
Critical
0
Units
30,470

Why these products were recalled

Due to a programmer software anomaly under very specific circumstance when executing a pacing capture Decrement Test in-clinic on implantable cardioverter and cardiac resynchronization therapy defibrillator devices, the programmer may continue to execute the Decrement Test instead of terminating the test and restoring the permanent programmed pacing parameters.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2022-04-27

    Cardiac Device Programmer Software May Fail to Terminate Test

    St. Jude Medical's Merlin 2 programmer software can fail to stop a clinical Decrement Test and restore pacing parameters during in-clinic testing, affecting 401 devices worldwide.

    Product
    Merlin 2 PCS MER3700 programmer Model MER3400 software
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2022-04-27

    Merlin PCS 3650 Programmer May Fail to Terminate Pacing Test Properly

    A software anomaly in the Merlin PCS 3650 programmer may cause an in-clinic pacing test to fail terminating and prevent restoration of the implanted device's normal settings. Approximately 30,069 affected devices have been distributed worldwide.

    Product
    Merlin PCS 3650 programmer Model 3330 software
    Category
    Distribution
    50 states