FDA (Devices) recall action 89810

Graphic Controls Acquisition Corporation recalled 6 products on 2022-04-20

The FDA (Devices) publishes a recall like this one as 6 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 6 pages.

Products
6
Announced
2022-04-20
Critical
0
Units
28,335

Why these products were recalled

Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–6 of 6

  • HighFDA (Devices)··2022-04-20

    CLARAVUE ECG Electrode May Fail to Deliver Effective Defibrillation

    CLARAVUE Pre-wired ECG electrodes may not meet electrical safety standards for defibrillation, potentially delivering an ineffective electrical charge during emergency use. This is a Class II recall affecting devices nationwide.

    Product
    CLARAVUE Pre-wired Electrode - IP SET 5V RT ADULT Model: 32028781 50604-US
    Category
    Distribution
    11 states
  • HighFDA (Devices)··2022-04-20

    ECG Electrode Set Fails Electrical Safety, May Prevent Effective Defibrillation

    Graphic Controls' CLARAVUE pediatric electrode set fails electrical safety standards. The device may not deliver adequate charge during defibrillation, potentially rendering emergency treatment ineffective.

    Product
    CLARAVUE Pre-wired Electrode IP-SET N3 - PEDIATRIC-US Model: 32028776 50401-US
    Category
    Distribution
    11 states
  • HighFDA (Devices)··2022-04-20

    Recalled CLARAVUE Pre-wired Electrode Set May Fail During Defibrillation

    Graphic Controls is recalling CLARAVUE Pre-wired Electrode sets because they fail to meet electrical safety specifications for defibrillators, potentially delivering inadequate charge during emergency treatment.

    Product
    CLARAVUE Pre-wired Electrode - IP SET M3 RT ADULT-US Model: 32028777 50600 US
    Category
    Distribution
    11 states
  • HighFDA (Devices)··2022-04-20

    Cardiac electrode fails electrical safety test; defibrillation may not work

    The CLARAVUE Pre-wired Electrode does not meet electrical safety standards for defibrillators. It may not deliver effective electrical charge during emergency defibrillation.

    Product
    CLARAVUE Pre-wired Electrode - IP SET 6V RT ADULT-US Model: 32028780 50603-US
    Category
    Distribution
    11 states
  • HighFDA (Devices)··2022-04-20

    Pre-wired ECG Electrode Set Fails Defibrillation Voltage Safety Standard

    Graphic Controls' CLARAVUE Pre-wired Electrode set does not meet required electrical specifications for defibrillation use. The device fails to withstand 5,000 volts without breakdown, potentially resulting in ineffective defibrillation.

    Product
    CLARAVUE Pre-wired Electrode - IP SET 12 RT ADULT-US Model: 32028779 50602-US
    Category
    Distribution
    11 states
  • HighFDA (Devices)··2022-04-20

    CLARAVUE Pre-wired Electrode fails to meet electrical safety specification

    Graphic Controls' CLARAVUE Pre-wired Electrode units fail to meet electrical safety specifications required for defibrillation. The device delivers an inefficient electrical charge and may result in ineffective defibrillation.

    Product
    CLARAVUE Pre-wired Electrode - IP SET M5 RT ADULT-US Model: 32028778 50601-US
    Category
    Distribution
    11 states