The Recall Desk

FDA (Devices) recall action 89784

B Braun Medical Inc recalled 2 products on 2022-04-20

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2022-04-20
Critical
0
Units
175

Why these products were recalled

Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–2 of 2

  • HighFDA (Devices)··2022-04-20

    Conductivity Sensors in Dialog Hemodialysis Systems May Crack and Leak

    B Braun Medical Inc is recalling conductivity sensors used in Dialog and Dialog+ hemodialysis systems due to potential cracks that could cause dialysate circuit leaks, affecting ultrafiltration during treatment.

    Product
    Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (BIC) - VERSION 2, Article Code Number 3456102A. These conductivity sensors may be used as spare parts for the following dialysis machines: 710200L (UDI 04046963686345), 710200S (UDI…
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-04-20

    Hemodialysis Conductivity Sensors recalled for potential cracks and dialysate leaks

    B Braun Medical Inc is recalling conductivity sensors used in Dialog hemodialysis systems due to potential cracking during use, which could cause leaks in the dialysate circuit and improper ultrafiltration. Affected patients should contact their healthcare provider.

    Product
    Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Article Code Number 3456103A. These conductivity sensors may be used as spare parts for the following dialysis machines: 710200L (UDI 04046963686345), 710200S (UDI…
    Category
    Distribution
    0 states