The Recall Desk

FDA (Devices) recall action 89544

Biomet, Inc. recalled 8 products on 2022-03-02

The FDA (Devices) publishes a recall like this one as 8 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 8 pages.

Products
8
Announced
2022-03-02
Critical
0
Units
96

Why these products were recalled

The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–8 of 8

  • HighFDA (Devices)··2022-03-02

    Biomet Shoulder Implant Stem Recalled for Inadequate Coating Adhesion

    Biomet is recalling a Comprehensive Shoulder System implant component due to a defect in the porous plasma coating. The coating may not have sufficient adhesion strength, compromising the implant's structural integrity.

    Product
    Comprehensive Shoulder System Primary Shoulder Stem, Micro Length, Porous Plasma, 14 MM, 55 MM Long Item Number: 113614
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-03-02

    Shoulder Implant Stem Coating Defect Due to Invalid Rework Process

    Biomet is recalling a shoulder implant stem due to a manufacturing defect in the porous plasma coating. Internal testing indicated the coating may not have sufficient adhesion strength after an unvalidated waterjet rework process.

    Product
    Comprehensive Shoulder System Primary Shoulder Stem, Micro Length, Porous Plasma, 11 MM, 55 MM Long Item Number: 113611
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-03-02

    Hip implant revision system recalled due to inadequate coating adhesion

    Biomet is recalling 19 units of the Arcos Modular Revision Hip System because an incorrect rework process may have weakened the coating's adhesion strength, posing a potential implant failure risk.

    Product
    Arcos Modular Revision Hip System, Standard Cone Prox Body, Porous Plasma,Size A, 60 MM Item Number: 11-301301
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-03-02

    Biomet Shoulder Implant Stem Recalled for Coating Adhesion Defect

    Biomet recalls 9 Comprehensive Shoulder System Mini Humeral Stems due to potential coating adhesion defects. An incorrect rework process may have compromised the porous plasma coating adhesion strength.

    Product
    Comprehensive Shoulder System Mini Humeral Stem, Porous Plasma, 13 MM, 83 MM Long Item Number: 113633
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-03-02

    Shoulder Implant Stem Recalled for Defective Coating Adhesion

    Biomet is recalling Comprehensive Shoulder System Mini Humeral Stems due to defective porous plasma coating adhesion. Internal testing indicated affected parts may lack sufficient coating strength.

    Product
    Comprehensive Shoulder System Mini Humeral Stem, Porous Plasma, 15 MM, 83 MM Long Item Number: 113635
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-03-02

    Biomet Shoulder Stem Implants Recalled for Coating Adhesion Defect

    Biomet is recalling certain shoulder stem implants due to potential insufficient adhesion strength in the porous plasma coating. An improper rework process may have compromised the coating's integrity.

    Product
    Comprehensive Shoulder System Primary Shoulder Stem, Micro Length, Porous Plasma, 18 MM, 55 MM Long Item Number: 113618
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-03-02

    Comprehensive Shoulder System Mini Humeral Stem Recalled for Coating Adhesion Defect

    Biomet recalls 11 units of its Comprehensive Shoulder System Mini Humeral Stem due to potential coating adhesion defects. An unvalidated rework process was applied to vendor-coated parts, risking implant failure.

    Product
    Comprehensive Shoulder System Mini Humeral Stem, Porous Plasma, 11 MM, 83 MM Long Item Number: 113631
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2022-03-02

    Biomet Shoulder Implant Stem Coating Adhesion Defect Recall

    Biomet is recalling shoulder implant stems due to improper rework of porous plasma coating, which may lack sufficient adhesion strength. Affected units may fail if the coating detaches.

    Product
    Comprehensive Shoulder System Mini Humeral Stem, Porous Plasma, 12 MM, 83 MM Long Item Number: 113632
    Category
    Distribution
    0 states