FDA (Devices) recall action 89276

Philips Respironics, Inc. recalled 2 products on 2022-02-02

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2022-02-02
Critical
0
Units
678

Why these products were recalled

Philips has identified a specific lot of non-conforming material manufactured by one of its suppliers. A Philips supplier incorrectly used polyester-based polyurethane (PE-PUR) sound abatement foam, a non-conforming material in the devices and repair kits. PE-PUR foam may degrade into particles which may enter the device s air pathway and be ingested or inhaled by the user. The foam degradation may be exacerbated by use of unapproved cleaning methods such as unapproved ozone cleaners, and the PE-PUR foam may off-gas certain chemicals during initial use of the device.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • SevereFDA (Devices)··2022-02-02

    Trilogy Evo Ventilator Repair Kit Recalled for Degrading Foam Material

    Philips Respironics recalled 205 Trilogy Evo ventilator repair kits containing non-conforming foam material that may degrade into particles potentially inhaled or ingested by users. The affected kits were distributed in the US and internationally.

    Product
    Trilogy Evo Repair Kit, RP-Trilogy Evo Muffler Assembly, Material Number 1135257 The Trilogy Evo ventilator provides continuous or intermittent positive pressure ventilation for the care of individuals who require mechanical ventilation. Trilogy Evo is intended for pediatric…
    Category
    Distribution
    17 states
  • SevereFDA (Devices)··2022-02-02

    Trilogy Evo Ventilator Recalled for Foam Degradation Risk

    Philips Respironics recalls 473 Trilogy Evo ventilators due to non-conforming foam that may degrade and release particles. Affected patients risk inhaling or ingesting foam particles and chemicals during device use.

    Product
    Trilogy Evo, Material Numbers DS2110X11B (USA) and KR2110X15B (Korea) The Trilogy Evo ventilator provides continuous or intermittent positive pressure ventilation for the care of individuals who require mechanical ventilation. Trilogy Evo is intended for pediatric through…
    Category
    Distribution
    17 states