FDA (Devices) recall action 88880

Abbott Laboratories recalled 2 products on 2021-12-01

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2021-12-01
Critical
0
Units
28,529

Why these products were recalled

Twelve software-related issues affecting software version 9.41 and earlier were identified. The issues include bypassing error codes, incorrect calibration, loss of module communication leading to overfilling wash buffer containers, missing QC flags, signal spikes, abnormal optics reads, and incorrect sample aspiration. The issues may lead to inaccurate results with an unknown bias direction and magnitude for assays measuring cardiac, metabolic (including diabetes), toxicology, prenatal, therapeutic drug monitoring, infectious disease, oncology, hepatic, pancreatic, hematological, endocrinolog…

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2021-12-01

    Abbott ARCHITECT Analyzer Software Issues May Cause Inaccurate Test Results

    Abbott is recalling approximately 20,994 ARCHITECT i1000SR, i2000SR, and i2000 automated analyzers due to 12 software-related issues in version 9.41 and earlier that may produce inaccurate diagnostic test results.

    Product
    ARCHITECT i1000SR REF 01L86-01/1L86 and 1L87; ARCHITECT i2000SR REF 03M74-02/3M74; ARCHITECT i2000 REF 08C89-01/1G17 and 8C89
    Category
    Distribution
    50 states
  • HighFDA (Devices)··2021-12-01

    Clinical Chemistry Analyzer Software Defects May Produce Inaccurate Results

    Abbott ARCHITECT clinical chemistry analyzers with software 9.41 and earlier contain defects affecting calibration, sample handling, and quality control that may produce inaccurate diagnostic results.

    Product
    ARCHITECT c4000 REF 02P24-01/2P24, 1P86, 1R24, and 1R25; ARCHITECT c8000 REF 01G06-11/1G06; ARCHITECT c16000 REF 03L77-01/3L77;
    Category
    Distribution
    50 states