Why these products were recalled

There is a potential that the length of bone screws identified on labeling may not correspond to the actual length of the screw included.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2021-10-27

    Orthopedic Bone Screws Recalled Due to Potential Length Mislabeling

    Limacorporate S.p.A is recalling 151 units of bone screws because the labeled length may not match the actual screw length included in the package.

    Product
    REF 8420.15.020, Bone Screw/Vite - Ti6AI4V, Self Tapping/Autofilettante, Dia=6.5mm, h=25mm, STERILE R, 2100099, UDI: (01)08033390018852
    Category
    Distribution
    13 states
  • HighFDA (Devices)··2021-10-27

    Bone Screw Labeling Error: Actual Length May Not Match Package Label

    Limacorporate is recalling bone screws (REF 8420.15.010) because the actual length of the screws may not correspond to the length stated on the package label, potentially affecting proper surgical placement and implant stability.

    Product
    REF 8420.15.010, Bone Screw/Vite - Ti6AI4V, Self Tapping/Autofilettante, Dia=6.5mm, h=20mm, STERILE R, 2100099, UDI: (01)08033390018845
    Category
    Distribution
    13 states