Biodex Medical Systems, Inc. recalled 2 products on 2021-10-06

The FDA (Devices) publishes a recall like this one as 2 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 2 pages.

Products
2
Announced
2021-10-06
Critical
0
Units
1,258

Why these products were recalled

When deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may then have incorrect dial settings and may lead to an incorrect dose measurement and/or misadministration

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–2 of 2

  • HighFDA (Devices)··2021-10-06

    Dose Calibrator Software Bug May Cause Incorrect Isotope Measurements

    AtomLab 500Plus Dose Calibrator software versions 2.0.00–2.0.08 contain a bug that fails to remove dial settings when deleting custom isotopes, potentially causing incorrect dose calibrations and misadministration.

    Product
    AtomLab 500Plus Dose CalibratorSoftware 2.0.00 through Revision 2.0.08, inclusive Model: 086-332
    Category
    Distribution
    1 state
  • HighFDA (Devices)··2021-10-06

    Atomlab 500 Dose Calibrator Software Recall for Incorrect Dose Measurement

    Biodex Medical Systems is recalling Atomlab 500 Dose Calibrator Software versions 2.0.00 through 2.0.08 due to a software defect that can cause incorrect dose measurements and potential misadministration of radioactive material.

    Product
    Atomlab 500 Dose Calibrator Software 2.0.00 through Revision 2.0.08, inclusive Model: 086-330.
    Category
    Distribution
    1 state