FDA (Devices) recall action 88548

Hill-Rom, Inc. recalled 5 products on 2021-10-06

The FDA (Devices) publishes a recall like this one as 5 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 5 pages.

Products
5
Announced
2021-10-06
Critical
0
Units
700

Why these products were recalled

Stretch Leveller has the potential to lose the function of maintaining a patient in the horizontal position when lifting and in motion. When this function is lost the patient can slide backwards and potentially may fall out of the Oct lift sheet back onto the surface it is being lifted from.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

States named across these recalls

1–5 of 5

  • HighFDA (Devices)··2021-10-06

    Patient lift devices may lose ability to maintain horizontal positioning during transfers

    Hill-Rom LikoStretch Mod 600 IC patient lift devices may fail to maintain patients in a horizontal position while lifting, potentially causing patients to slide or fall.

    Product
    LikoStretch Mod 600 IC Model Number: 3156065 LikoStretch Mod 600 IC, Wide Model Number: 3156065B (obsolete since 12JUN2020) - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient.
    Category
    Distribution
    14 states
  • HighFDA (Devices)··2021-10-06

    Patient lift device may lose horizontal positioning control, creating fall risk

    The OctoStretch with Stretch Leveller may lose its ability to keep patients horizontal during lifting, risking patient falls. 440 units were distributed worldwide.

    Product
    OctoStretch with Stretch Leveller - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient. Model Number: 3156056
    Category
    Distribution
    14 states
  • HighFDA (Devices)··2021-10-06

    Patient Lift Device Stretch Leveller May Lose Horizontal Position Control

    Hill-Rom LikoStretch 1900 patient lift devices may lose their ability to maintain horizontal position during lifting, risking patient falls. Affected devices can cause patients to slide out of the lift sheet.

    Product
    LikoStretch 1900 Model Number: 3156051 - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient.
    Category
    Distribution
    14 states
  • HighFDA (Devices)··2021-10-06

    Medical Device: FlexoStretch Patient Lift Leveller May Lose Function During Patient Transfer

    Hill-Rom's FlexoStretch Model 3156057 patient lift device may lose the ability to maintain horizontal positioning during patient transfer, risking patient falls. The FDA Class II recall affects 129 units distributed worldwide.

    Product
    FlexoStretch Model Number: 3156057 - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient.
    Category
    Distribution
    14 states
  • HighFDA (Devices)··2021-10-06

    Stretch Leveller Model 3156200 May Lose Horizontal Position Control

    Hill-Rom's Stretch Leveller Model 3156200 may lose the ability to maintain patients in a horizontal position during lifting, potentially causing patients to slide backward or fall from the lift sheet.

    Product
    Stretch Leveller Model Number: 3156200 - The combination of this products allows the user to easily adjust/control the center of gravity during lift of patient.
    Category
    Distribution
    14 states