FDA (Devices) recall action 88531

B. Braun Medical, Inc. recalled 17 products on 2021-10-06

The FDA (Devices) publishes a recall like this one as 17 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 17 pages.

Products
17
Announced
2021-10-06
Critical
0
Units
1,717,730

Why these products were recalled

Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing and bloodstream infections.

Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing and bloodstream infections.

The agency’s own wording, quoted from the enforcement report.

The unit total adds up the quantities the agency stated for each product. Where a product record gave no figure, or gave a weight rather than a count, it contributes nothing — so the real total is at least this, never less.

1–17 of 17

  • HighFDA (Devices)··2021-10-06

    Outlook Pump Add-On Set Recalled for Potential Leakage and Under-Delivery

    B. Braun Medical is recalling the Outlook Pump Add-On Set due to potential leakage that may delay medication delivery and cause incomplete dosing. The affected units could increase the risk of bloodstream infections.

    Product
    Outlook Pump Add-On Set-single use, for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: V7400
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    B. Braun Pump Set Cassette Leakage Risks Medication Delivery

    B. Braun Medical recalls 13,320 pump sets due to cassette leakage that may delay medication administration and cause incomplete dosing or bloodstream infections.

    Product
    PUMP SET, 15 DR, 3 CARESITE, 127 in.single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 480239
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    Outlook Pump Measured Volume Solution Set recalled for cassette leakage

    B. Braun is recalling 940 units of Outlook Pump Measured Volume Solution Sets due to potential leakage in the cassette. The leakage could delay medication delivery, result in incomplete dosing, or lead to bloodstream infections.

    Product
    Outlook Pump Measured Volume Solution Set/ w/ Auto Shut-Off. Single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 375118
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    B. Braun Infusion Pump Sets Recalled for Leakage and Infection Risk

    B. Braun Medical's infusion pump sets may leak, potentially delaying medication delivery and causing incomplete dosing. The defect poses a risk of bloodstream infections.

    Product
    15 drops/ml, Priming Volume: 27ml, Length: 129 in. Single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluid Material ID: 354212
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    Outlook Pump Primary Administration Set Recalled for Potential Leakage and Medication Underdosing

    B. Braun Medical is recalling the Outlook Pump Primary Administration Set due to potential leakage that could delay medication delivery and cause incomplete dosing of critical fluids in epidural administration.

    Product
    Outlook Pump Primary Administration Set w/ micro Tubing for Epidural Administration-single use, for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: V7453
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    B. Braun Outlook Pump Set 3 Recalled for Leakage and Medication Delivery Delays

    B. Braun Medical is recalling 768 units of Outlook Pump Set 3 infusion sets due to potential leakage that could delay or prevent complete medication delivery and increase risk of bloodstream infection.

    Product
    OUTLOOK PUMP SET 3 CARESITE LADS 133 IN.-single use, for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 490564
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    B. Braun Outlook IV Pump Sets Recalled Due to Leakage Risk

    B. Braun Medical is recalling Outlook pump IV sets used with specific infusion pumps due to leakage in the cassette portion. The defect may prevent proper medication delivery and pose a risk of bloodstream infection.

    Product
    OUTLOOK PUMP IV SET W/UNIVERSAL SPIKE-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 474004
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    B. Braun Outlook Pump Set Recall: Risk of Leakage and Medication Under-Delivery

    B. Braun is recalling 1,752 Y-Type Blood Outlook Pump Set units due to leakage in the cassette portion. The defect may cause delays in medication administration, incomplete dosing, or bloodstream infections.

    Product
    Y-Type Blood, Outlook Pump Set, Caresite-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 477005
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    B. Braun Infusion Pump Sets Recalled Due to Cassette Leakage Risk

    B. Braun Medical is recalling 1,960 infusion pump sets used with Outlook Safety and Horizon NXT systems due to leakage in the cassette portion, which may delay medication delivery and increase bloodstream infection risk.

    Product
    60 drops/ml, Priming volume: 25 ml, Length: 120 in.in-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 354215
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    Blood Administration Pump Set Recalled Due to Leakage Risk

    B. Braun Medical is recalling Outlook Pump Blood Administration Sets due to leakage in the pump cassette that could delay or prevent complete medication delivery and increase infection risk.

    Product
    Outlook Pump Blood Administration Set with SAFELINE Injection Site-single use, for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: NF3120
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    Infusion Pump Sets Recalled for Leakage and Medication Under-Delivery Risk

    B. Braun Medical is recalling infusion pump sets that may leak, potentially causing delayed medication delivery, incomplete dosing, and bloodstream infections.

    Product
    10 drops/ml, Priming volume: 46ml, Length: 130 in.-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 354216
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    Outlook Pump Set Infusion System Recalled for Leakage and Medication Delivery Risk

    B. Braun is recalling 7,248 units of Outlook Pump Set that may leak or fail to deliver medications properly, potentially causing bloodstream infections.

    Product
    OUTLOOK PUMP SET, 3 CARESITE LADS, 115 IN-single use, for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 490361
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    B. Braun Outlook Pump Sets recalled for potential leakage and infection risk

    B. Braun is recalling 585,576 Outlook Pump Sets due to potential leakage in the cassette portion. The defect may cause medication under-delivery and increase bloodstream infection risk.

    Product
    Outlook Pump Sets w/3 CARESITE Needle-Free Injection Sites: single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Material ID: 352413
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    Outlook pump extension sets recalled due to leakage and dosing errors

    Outlook infusion pump extension sets may leak, potentially delaying medication delivery and causing bloodstream infections.

    Product
    ET82HP OUTLOOK PUMP EXTENSION SET-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 470024
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    Outlook Pump IV Sets Recalled for Leakage and Infection Risk

    B. Braun Medical, Inc. is recalling approximately 44,866 Outlook Pump IV sets due to leakage of the pump set that could delay medication administration and increase the risk of bloodstream infections. The defect may result in under-delivery or incomplete medication dosing.

    Product
    Outlook¿ Pump 15 Drop 0.2 Micron Filtered IV Set with 2 CARESITE¿ Y-Sites: single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 354214
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    Outlook V1500HP Pump Set Recalled for Cassette Leakage

    The Outlook V1500HP Pump Set is recalled due to cassette leakage that could delay medication delivery, reduce dosing accuracy, and cause bloodstream infections. The recall involves 12,000 units distributed in the US and Canada.

    Product
    V1500HP OUTLOOK PUMP SET 118-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 470025
    Category
    Distribution
    0 states
  • HighFDA (Devices)··2021-10-06

    Outlook Pump IV Set Recalled Due to Cassette Leakage and Infection Risk

    B. Braun Medical is recalling Outlook Pump IV Sets due to cassette leakage that may delay medication delivery and cause incomplete dosing or bloodstream infections.

    Product
    OUTLOOK PUMP IV SET, 134 IN.-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids Ref: 477009
    Category
    Distribution
    0 states