Medical Device Recall: TruMatch CT Cut Guide Kit Software Misalignment Error
DePuy Orthopaedics is recalling TruMatch CT Cut Guide Kits due to a software coding error in the Fast3D Segmentation software that may cause improper alignment of anatomic images, potentially resulting in limb malalignment during surgery.
- Product
- TRUMATCH CT CUT GUIDE KIT L- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number:
- Category
- Distribution
- 10 states