The Recall Desk

FDA (Devices) recall action 88522

DePuy Orthopaedics, Inc. recalled 7 products on 2021-09-08

The FDA (Devices) publishes a recall like this one as 7 separate product records. They share a firm, a date and a cause; this page is all of them in one place, so the scale of the action is visible rather than spread across 7 pages.

Products
7
Announced
2021-09-08
Critical
0
Units
—

Why these products were recalled

Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.

The agency’s own wording, quoted from the enforcement report.

States named across these recalls

1–7 of 7

  • HighFDA (Devices)··2021-09-08

    DePuy TruMatch CT Cut Guide Kit Recalled for Software Alignment Error

    DePuy's TruMatch CT Cut Guide Kit is being recalled due to a software coding error that may cause misalignment of preoperative surgical positioning scans, potentially resulting in improper limb alignment during joint replacement surgery.

    Product
    TRUMATCH CT CUT GUIDE KIT R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number…
    Category
    Distribution
    10 states
  • HighFDA (Devices)··2021-09-08

    Surgical instrument software error may cause limb misalignment

    DePuy Orthopaedics' TruMatch CT Pin Guide Kit has a software coding error that may misalign bone landmarks during joint replacement surgery, potentially causing limb malalignment. No injuries have been reported.

    Product
    TRUMATCH CT PIN GUIDE KIT R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number…
    Category
    Distribution
    10 states
  • HighFDA (Devices)··2021-09-08

    CT Pin Guide Instruments Recalled for Software Imaging Alignment Error

    The TruMatch CT Pin Guide Kit is being recalled due to a software error that may misalign surgical guidance images, potentially affecting limb positioning during hip, knee, and ankle surgery. No injuries have been reported.

    Product
    TRUMATCH CT PIN GUIDE KIT L- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part…
    Category
    Distribution
    10 states
  • HighFDA (Devices)··2021-09-08

    Surgical alignment software error recalled in patient-specific orthopedic instruments

    DePuy Orthopaedics recalls TruMatch surgical instruments due to a software error in the scanning alignment tool that may cause incorrect limb positioning during surgery.

    Product
    TRUMATCH CT PIN GUIDE FEM R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number…
    Category
    Distribution
    10 states
  • HighFDA (Devices)··2021-09-08

    TruMatch Surgical Instruments Recalled for Software Alignment Error

    DePuy Orthopaedics recalls TruMatch patient-specific surgical instruments due to a Fast3D software error that may misalign scanning images, potentially causing limb malalignment during surgery.

    Product
    TRUMATCH CT FEM CUT TIB PIN R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part…
    Category
    Distribution
    10 states
  • HighFDA (Devices)··2021-09-08

    TruMatch CT Surgical Guidance Software Misalignment in Joint Replacement Instruments

    DePuy Orthopaedics' TruMatch CT surgical instrument system has a software coding error that may prevent proper alignment of scanning images, potentially causing limb malalignment during joint replacement surgery.

    Product
    TRUMATCH CT CUT GUIDE FEM L- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number…
    Category
    Distribution
    10 states
  • HighFDA (Devices)··2021-09-08

    Medical Device Recall: TruMatch CT Cut Guide Kit Software Misalignment Error

    DePuy Orthopaedics is recalling TruMatch CT Cut Guide Kits due to a software coding error in the Fast3D Segmentation software that may cause improper alignment of anatomic images, potentially resulting in limb malalignment during surgery.

    Product
    TRUMATCH CT CUT GUIDE KIT L- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number…
    Category
    Distribution
    10 states